Anesthesia workstation performance in pediatric ventilation – Part 1: Precision of ventilation.
Authors:
- 13 May 2026
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesiology
- Vol. 145 (4) , 942-953
Abstract
Background.: Mechanical ventilation of children sets the highest demands on the performance of anesthesia workstations. Different ventilator technologies are available, each representing unique pneumatic characteristics. We hypothesized that precision of delivered ventilation parameters depends on the ventilator type. Methods.: Six modern anesthesia workstations (GE HealthCare Carestation 650; Mindray Bio-Medical Electronics Co. Ltd. Wato EX-65 Pro and A9; Dräger Medical Atlan A350 and Perseus A500; Getinge Flow-c) were tested in physical models simulating the respiratory systems of a preterm baby, a neonate, an infant and a toddler. Volume-controlled and pressure-controlled ventilation were investigated with and without airway leakage. Delivered tidal volume, peak inspiratory pressure, positive end-expiratory pressure and inspiratory to expiratory ratio were compared to the set values. Data are presented descriptively, a difference ≥ 10 % between delivered and the set values was considered clinically relevant. Results.: Deviations of the delivered volume from the set tidal volume during volume-controlled ventilation ranged between -19.5 ± 0.6 % and +35.1 ± 3.0 % in the preterm baby model, but decreased with increasing tidal volume. Positive end-expiratory pressure and peak inspiratory pressure during pressure-controlled ventilation did not differ relevantly from the set values in any device or lung model. Expiratory ratios varied widely in most devices leading to relevant differences in tidal volume during pressure-controlled ventilation. With leakage, tidal volume was on average 7 % lower during volume-controlled ventilation but 2 % lower during pressure-controlled ventilation. Conclusions.: Precision of ventilation with modern anesthesia workstations varies depending on the ventilator type, ventilation mode and the set parameters. Pressure-controlled ventilation is advantageous in the six workstations tested considering the good agreement between the applied and the set parameters and leakage compensation. Our results imply precision superiority of pressure-controlled ventilation over volume-controlled ventilation in preterm babies and neonates, independent from the ventilator technology in use.Keywords
This publication has 29 references indexed in Scilit:
- Design and development of an affordable multi-mode small animal ventilatorScientific Reports, 2025
- Calculating intrinsic positive end‐expiratory pressure from end‐expiratory flow in mechanically ventilated children—A study in physical models of the pediatric respiratory systemPediatric Pulmonology, 2023
- Trigger performance of five pediatric home ventilators and one ICU ventilator depending on circuit type and system leak in a physical model of the lungPediatric Pulmonology, 2021
- Accuracy of volume and pressure delivery by mechanical ventilators in use in neonatal intensive care units: A quality control studyPediatric Pulmonology, 2020
- Recommendations for mechanical ventilation of critically ill children from the Paediatric Mechanical Ventilation Consensus Conference (PEMVECC)Intensive Care Medicine, 2017
- Tidal Volume Delivery during the Anesthetic Management of Neonates Is VariableThe Journal of Pediatrics, 2017
- FLOW-i ventilator performance in the presence of a circle system leakJournal of Clinical Monitoring and Computing, 2016
- Intraoperative mechanical ventilation for the pediatric patientBest Practice & Research Clinical Anaesthesiology, 2015
- Flow measurement in mechanical ventilation: A reviewMedical Engineering & Physics, 2015
- Elevated inspiratory CO2 concentrations caused by the design of an anesthesia machineActa Anaesthesiologica Scandinavica, 2003