Measurements of in-air spot size of pencil proton beam for various air gaps in conjunction with a range shifter on a ProteusPLUS PBS dedicated machine and comparison to the proton dose calculation algorithms
Authors:
- 20 June 2019
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 42 (3) , 853-862
Abstract
The purpose of this study is to (i) investigate the impact of various air gaps in conjunction with a range shifter of 7.5 cm water-equivalent-thickness (WET) on in-air spot size of a pencil proton beam at the isocenter and off-axis points, and (ii) compare the treatment planning system (TPS) calculated spot sizes against the measured spot sizes. A scintillation detector has been utilized to measure the in-air spot sizes at the isocenter. The air gap was varied from 0 to 35 cm at an increment of 5 cm. For each air gap, a single spot pencil proton beam of various energies (110–225 MeV) was delivered to the scintillation detector. By mimicking the experimental setup in RayStation TPS, proton dose calculations were performed using pencil beam (RS-PB) and Monte Carlo (RS-MC) dose calculation algorithms. The calculated spot sizes (RS-PB and RS-MC) were then compared against the measured spot sizes. For a comparative purpose, the spot sizes of each measured energy for different air gaps of (5–35 cm) were compared against that of 0 cm air gap. The results of the 5 cm air gap showed an increase in spot size by ≤ 0.6 mm for all energies. For the largest air gap (35 cm) in the current study, the spot size increased by 3.0 mm for the highest energy (225 MeV) and by 9.2 mm for the lowest energy (110 MeV). For the 0 cm air gap, the agreement between the TPS-calculated (RS-PB and RS-MC) and measured spot sizes were within ± 0.1 mm. For the 35 cm air gap, the RS-PB overpredicted spot sizes by 0.3–0.8 mm, whereas the RS-MC computed spot sizes were within ± 0.3 mm of measured spot sizes. In conclusion, spot size increment is dependent on the energy and air gap. The increase in spot size was more pronounced at lower energies ( < 150 MeV) for all air gaps. The comparison between the TPS calculated and measured spot sizes showed that the RS-MC is more accurate (within ± 0.3 mm), whereas the RS-PB overpredicted (up to 0.8 mm) the spot sizes when a range shifter (7.5 cm WET) and large air gaps are encountered in the proton beam path.Keywords
This publication has 16 references indexed in Scilit:
- Quantifying the effect of air gap, depth, and range shifter thickness on TPS dosimetric accuracy in superficial PBS proton therapyJournal of Applied Clinical Medical Physics, 2017
- Commissioning of the world's first compact pencil‐beam scanning proton therapy systemJournal of Applied Clinical Medical Physics, 2017
- Comparison of multi‐institutional Varian ProBeam pencil beam scanning proton beam commissioning dataJournal of Applied Clinical Medical Physics, 2017
- A benchmarking method to evaluate the accuracy of a commercial proton monte carlo pencil beam scanning treatment planning systemJournal of Applied Clinical Medical Physics, 2017
- Clinical Commissioning of a Pencil Beam Scanning Treatment Planning System for Proton TherapyInternational Journal of Particle Therapy, 2016
- Development and Clinical Implementation of a Universal Bolus to Maintain Spot Size During Delivery of Base of Skull Pencil Beam Scanning Proton TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2014
- Adjustment of the lateral and longitudinal size of scanned proton beam spots using a pre-absorber to optimize penumbrae and delivery efficiencyPhysics in Medicine & Biology, 2010
- Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas M.D. Anderson Cancer Center, Proton Therapy Center, HoustonMedical Physics, 2009
- Density heterogeneities and the influence of multiple Coulomb and nuclear scatterings on the Bragg peak distal edge of proton therapy beamsPhysics in Medicine & Biology, 2008
- Multiple Coulomb scattering of 160 MeV protonsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1993