Radiation dose from 3D rotational X-ray imaging: organ and effective dose with conversion factors
- 16 September 2011
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 150 (1) , 50-54
Abstract
The purpose of this study was to measure organ doses and the effective dose (ED) using a three-dimensional rotational X-ray (3D-RX) system and to determine the ED conversion factor from the dose area product (DAP) for skull, spine and biliary protocols. A commercial 3D-RX imaging system was used to simulate the protocols with the adult female anthropomorphic phantom. Twenty MOSFET detectors were used to measure the absorbed doses at various organ locations. The ED was calculated for each protocol and the corresponding DAP was obtained. The skin dose was the highest for all the protocols. The second highest organ doses were those of the brain for the skull, the intestine for the spine and the kidney for the biliary protocol. The ED was 0.4–0.9, 4.2–8.4 and 3.2–4.6 mSv, and the ED conversion factor was 0.06–0.09, 0.18–0.31 and 0.13–0.23 mSv Gy−1 cm−2 for each protocol, respectively. This data may be used to estimate the patient ED for those protocols in the 3D-RX.Keywords
This publication has 13 references indexed in Scilit:
- Effective dose analysis of three-dimensional rotational angiography during catheter ablation proceduresPhysics in Medicine & Biology, 2010
- Three-dimensional cardiac rotational angiography: effective radiation dose and image quality implicationsEP Europace, 2009
- Patient radiation exposure during diagnostic and therapeutic interventional neuroradiology proceduresJournal of NeuroInterventional Surgery, 2009
- Comparison of Patient Dose in Two-Dimensional Carotid Arteriography and Three-Dimensional Rotational AngiographyCardioVascular and Interventional Radiology, 2007
- Patient dose from 3D rotational neurovascular studiesBritish Journal of Radiology, 2007
- Patient and Staff Dosimetry in VertebroplastySpine, 2006
- Three-Dimensional Rotational Angiography of Neurovascular Lesions in Pediatric PatientsAmerican Journal of Roentgenology, 2006
- Three-dimensional Rotational Angiography: Introduction of an Adjunctive Tool for Successful Transarterial ChemoembolizationJournal of Vascular and Interventional Radiology, 2005
- 3-Dimensional rotational X-ray imaging, 3D-RX: image quality and patient dose simulation for optimisation studiesRadiation Protection Dosimetry, 2005
- Dose Evaluation and Effective Dose Estimation in Radiological Studies of Paranasal SinusesRadiation Protection Dosimetry, 1999