Fatigue testing of a NiTi rotary instrument. Part 2: fractographic analysis
Authors:
- 10 July 2007
- journal article
- Published by Wiley in International Endodontic Journal
- Vol. 40 (8) , 619-625
Abstract
Aim To examine the topographic features of the fracture surface of a NiTi instrument after fatigue failure, and to correlate the measurements of some features with the cyclic load.Methodology A total of 212 ProFile rotary instruments were subjected to a rotational‐bending test at various curvatures until broken. The fracture surface of all fragments was examined by SEM to identify the crack origins. The crack radius, i.e. extent of the fatigue‐crack growth towards the centroid of the cross‐section, was also measured, and correlated with the strain amplitude for each instrument.Results All fracture surfaces revealed the presence of one or more crack origins, a region occupied by microscopic striations, and an area with microscopic dimples. The number of specimens showing multiple crack origins was significantly greater in the group fatigued under water than in air (P < 0.05). A linear relationship between the reciprocal of the square root of the crack radius and the strain amplitude was discernible (P < 0.001), the slopes of which were not significantly different for instruments fatigued in air and water.Conclusions The fractographic appearance of NiTi engine‐files that had failed because of fatigue is typical of that for other metals. The fatigue behaviour of NiTi instruments is adversely affected by water, not only for the low‐cycle fatigue life, but also the number of crack origins. There appears to be a critical extent of crack propagation for various strain amplitudes leading to final rupture (akin to the Griffith's criterion for brittle materials).Keywords
This publication has 13 references indexed in Scilit:
- Fatigue testing of a NiTi rotary instrument. Part 1: strain–life relationshipInternational Endodontic Journal, 2007
- Failure mechanism of ProTaper Ni–Ti rotary instruments during clinical use: fractographic analysisInternational Endodontic Journal, 2006
- Defects in ProTaper S1 instruments after clinical use: fractographic examinationInternational Endodontic Journal, 2005
- Decrease in the fatigue resistance of nickel-titanium rotary instruments after clinical use in curved root canalsOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2005
- Defects in ProTaper S1 instruments after clinical use: longitudinal examinationInternational Endodontic Journal, 2005
- Factors Influencing Defects of Rotary Nickel-Titanium Endodontic Instruments After Clinical UseJournal of Endodontics, 2004
- Comparative investigation of two rotary nickel–titanium instruments: ProTaper versus RaCe. Part 2. Cleaning effectiveness and shaping ability in severely curved root canals of extracted teethInternational Endodontic Journal, 2004
- Evaluation of Single-use Rotary Nickel-titanium InstrumentsJournal of Endodontics, 2003
- The Effect of Rotational Speed and the Curvature of Root Canals on the Breakage of Rotary Endodontic InstrumentsJournal of Endodontics, 2002
- Defects in Rotary Nickel-Titanium Files After Clinical UseJournal of Endodontics, 2000