Accurate field mapping in the presence of B0 inhomogeneities, applied to MR thermometry
- 27 June 2014
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 73 (6) , 2142-2151
Abstract
Purpose: To describe how B0 inhomogeneities can cause errors in proton resonance frequency (PRF) shift thermometry, and to correct for these errors.Methods: With PRF thermometry, measured phase shifts are converted into temperature measurements through the use of a scaling factor proportional to the echo time, TE. However, B0 inhomogeneities can deform, spread, and translate MR echoes, potentially making the “true” echo time vary spatially within the imaged object and take on values that differ from the prescribed TE value. Acquisition and reconstruction methods able to avoid or correct for such errors are presented.Results: Tests were performed in a gel phantom during sonication, and temperature measurements were made with proper shimming as well as with intentionally introduced B0 inhomogeneities. Errors caused by B0 inhomogeneities were observed, described, and corrected by the proposed methods. No statistical difference was found between the corrected results and the reference results obtained with proper shimming, while errors by more than 10% in temperature elevation were corrected for. The approach was also applied to an abdominal in vivo dataset.Conclusion: Field variations induce errors in measured field values, which can be detected and corrected. The approach was validated for a PRF thermometry application. Magn Reson Med 73:2142–2151, 2015.Keywords
Funding Information
- NIH (R01CA149342, P41EB015898, R01EB010195)
This publication has 33 references indexed in Scilit:
- Multipathway sequences for MR thermometryMagnetic Resonance in Medicine, 2011
- Temperature‐induced tissue susceptibility changes lead to significant temperature errors in PRFS‐based MR thermometry during thermal interventionsMagnetic Resonance in Medicine, 2010
- MR thermometryJournal of Magnetic Resonance Imaging, 2008
- Real-Time Monitoring of Hepatocellular Carcinoma Radiofrequency Ablation by Quantitative Temperature MRISeminars in Liver Disease, 2006
- Application of k-space energy spectrum analysis to susceptibility field mapping and distortion correction in gradient-echo EPINeuroImage, 2006
- Real‐time monitoring of radiofrequency ablation of rabbit liver by respiratory‐gated quantitative temperature MRIJournal of Magnetic Resonance Imaging, 2006
- In Vivo Temperature Mapping of Prostate during Treatment with TherMatrx TMx-2000 Device: Heat Field and MRI Determinations of Necrotic LesionsJournal of Endourology, 2005
- A precise and fast temperature mapping using water proton chemical shiftMagnetic Resonance in Medicine, 1995
- Noninvasive MRI Thermometry with the Proton Resonance Frequency (PRF) Method: In Vivo Results in Human MuscleMagnetic Resonance in Medicine, 1995
- Reduction of susceptibility artifact in gradient‐echo imagingMagnetic Resonance in Medicine, 1992