Jamming of 25‐gauge instruments in the cannula during vitrectomy for vitreous haemorrhage
Authors:
Open Access
- 1 March 2008
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 86 (2) , 160-164
Abstract
: Purpose: To report the jamming of 25‐gauge instruments in the cannula during vitreous surgery for non‐clearing vitreous haemorrhage.Methods: Forty‐five eyes underwent vitrectomy with 25‐gauge instruments for non‐clearing vitreous haemorrhage (VH group). The incidence of 25‐gauge instruments jamming in the cannula was determined retrospectively and compared with that in 112 eyes that underwent vitrectomy for epiretinal membrane (ERM group), also using 25‐gauge instruments.Results: The 25‐gauge vitreous cutter or light pipe became jammed in the cannula in three eyes (7%) in the VH group and the instrument locked inside the cannula had to be removed with the cannula. None of the 25‐gauge instruments in the ERM group jammed (p = 0.022, Fisher's exact probability test). Two of three eyes developed giant retinal breaks near the sclerotomy but no retinal break related to the sclerotomy was detected in the ERM group. Examination of the cutter revealed blood trapped between the cutter and the cannula.Conclusions: Twenty‐five gauge instruments may become jammed in the cannula in eyes with non‐clearing vitreous haemorrhage. Clinicians should be aware of this surgical complication when 25‐gauge instruments are used in vitreous haemorrhage.Keywords
This publication has 12 references indexed in Scilit:
- Entry‐site‐related retinal detachment after pars plana vitrectomyActa Ophthalmologica Scandinavica, 2007
- 25-gauge, sutureless vitrectomy and standard 20-gauge pars plana vitrectomy in idiopathic epiretinal membrane surgery: a comparative pilot studyGraefe's Archive for Clinical and Experimental Ophthalmology, 2006
- Sutureless Vitreoretinal Surgery Using a New 25-Gauge Transconjunctival SystemOphthalmic surgery, lasers & imaging retina, 2006
- Expanded Indications for 25-Gauge Transconjunctival VitrectomyJapanese Journal of Ophthalmology, 2005
- Longer-term Outcomes of Transconjunctival Sutureless 25-gauge VitrectomyAmerican Journal of Ophthalmology, 2005
- Outcomes of 140 Consecutive Cases of 25-Gauge Transconjunctival Surgery for Posterior Segment DiseaseOphthalmology, 2005
- Intraoperative breakage of a 25-gauge vitreous cutterAmerican Journal of Ophthalmology, 2004
- A new 25-gauge instrument system for transconjunctival sutureless vitrectomy surgery1 1The new Transconjunctival Sutureless Vitrectomy System is disclosed to Bausch and Lomb Surgical, St. Louis, MO. The Microsurgery Advanced Design Laboratory (MADLAB) may receive royalties related to the sale of this and other instruments mentioned in the article.Ophthalmology, 2002
- Initial experience using the transconjunctival sutureless vitrectomy system for vitreoretinal surgery1 1The new Transconjunctival Sutureless Vitrectomy System is disclosed to Bausch & Lomb Surgical, St. Louis, MO. The Microsurgery Advanced Design Laboratory (MADLAB) may receive royalties related to the sale of this and other instruments mentioned in the article.Ophthalmology, 2002
- Pars plana incisions of four patients: histopathology and electron microscopy.British Journal of Ophthalmology, 1995