The principle of low frictional torque in the Charnley total hip replacement
Authors:
Open Access
- 1 July 2009
- journal article
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 91-B (7) , 855-858
Abstract
The design of the Charnley total hip replacement follows the principle of low frictional torque. It is based on the largest possible difference between the radius of the femoral head and that of the outer aspect of the acetabular component. The aim is to protect the bone-cement interface by movement taking place at the smaller radius, the articulation. This is achieved in clinical practice by a 22.225 mm diameter head articulating with a 40 mm or 43 mm diameter acetabular component of ultra-high molecular weight polyethylene. We compared the incidence of aseptic loosening of acetabular components with an outer diameter of 40 mm and 43 mm at comparable depths of penetration with a mean follow-up of 17 years (1 to 40). In cases with no measurable wear none of the acetabular components were loose. With increasing acetabular penetration there was an increased incidence of aseptic loosening which reflected the difference in the external radii, with 1.5% at 1 mm, 8.8% at 2 mm, 9.7% at 3 mm and 9.6% at 4 mm of penetration in favour of the larger 43 mm acetabular component. Our findings support the Charnley principle of low frictional torque. The level of the benefit is in keeping with the predicted values.Keywords
This publication has 23 references indexed in Scilit:
- Effect of Reduced Diameter Neck Stem on Incidence of Radiographic Cup Loosening and Revisions in Charnley Low-Frictional Torque ArthroplastyThe Journal of Arthroplasty, 2009
- Low-friction arthroplasty of the hip using alumina ceramic and cross-linked polyethyleneThe Journal of Bone and Joint Surgery. British volume, 2005
- The Contribution of Frictional Torque to Loosening at the Cement-Bone Interface in Tharies Hip Replacements*Journal of Bone and Joint Surgery, 1996
- Charnley Low-Friction Arthroplasty of the HipClinical Orthopaedics and Related Research, 1993
- Socket Size and the Expected Quality of its Fixation in the Cemented Total Hip ArthroplastyProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1990
- Comparative experience with intramedullary and extramedullary alignment in total knee arthroplastyThe Journal of Arthroplasty, 1990
- Radiographic demarcation of the acetabular bone-cement interfaceThe Journal of Arthroplasty, 1990
- The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hipMedical & Biological Engineering & Computing, 1969
- ARTHROPLASTY OF THE HIP A New OperationThe Lancet, 1961
- Surgery of the Hip-jointBMJ, 1960