Simulation analysis of x-ray phase imaging with polycapillary optics

Authors:

Abstract
Currently, mammography utilizes subtle attenuation differences in the soft tissue of the breast to produce image contrast, but x-ray phase imaging can have inherently higher visibility. Rather than using gratings, which can be costly and require precision alignment, our x-ray phase system utilizes a low-cost wire mesh imaged with and without an object in place. Distortions of the mesh pattern can be analyzed to produce phase, dark field, and attenuation images, providing two additional channels of information. The use of a polycapillary optic to create a secondary source provides greater coherence. In order to generate test data for neural network training, and to optimize the system, a simulation has been developed which produces mesh and mesh-with-object images in a variety of conditions. Integrated phase retrieved from simulated images is in good agreement with theoretical values.

This publication has 24 references indexed in Scilit: