Design of Medical Ultrasound Probe: Measurement System for the Characterization of Reverberations

Authors:

Abstract
A measurement system which characterizes the reverberations occurring in medical ultrasound (US) images has been developed. Reverberations cause multiple copies of anatomical structure interfaces, degrading image quality, accuracy and leading to potential misdiagnosis. The system is suitable for characterizing the reverberations during design phase of US probes and for investigating the root causes of these distortions that influence the medical images. The system, based on non-destructive evaluation of the materials, is implemented by means of an ad hoc measurement chain, in which a silicone membrane specially designed to generate reverberations, replicates the same acoustic properties of human skin. The experimental setup based Time of Flight (ToF) of the echoes allows for detecting the reverberation source in the US probe acoustic stack. The proposed quantitative approach can help to predict US probe performances during design stage becoming very useful during imaging validation. Accurate description of the ultrasonic transmit-receive measurement system is discussed by taking into account different material, geometry and thickness of the layers involved in the US probe. Experimental results show that the implemented system can detect the root cause of reverberations by accurately quantifying the layer source providing imaging distortion.