Current Organizations / Institutes

The organizations/institutes from the scholar's latest article

Previous Organizations / Institutes

The organizations / institutes from the scholar's other articles in the past ten years


Francesco Bertocci has a total of 52 co-authors. Here are the top co-authors by number of shared publications:


Latest Publications

  • Open Access

    Reliability with multiple causes of failures: Modeling and practice through a case study on ultrasound probes for medical imaging

    by Rossella Berni, Francesco Bertocci, Alessandro Magrini, Nedka D. Nikiforova

    In this paper, we deal with statistical modeling and a related case study for reliability when multiple failure causes are present. At first, we present in detail two main approaches for competing risk modeling, e.g. the Cox Proportional Hazards model, and the Fine & Gray model. In both models, we consider the inclusion of random effects, a no-trivial issue in this context, especially from the practical point of view. Following, we deal with advanced statistical models to compare the causes of failure, providing extremely useful information for production managers. To perform a useful study for practitioners, statistical modeling is illustrated through an empirical example related to ultrasound probes for medical imaging. The main theory is briefly presented comprehensively, while particular emphasis is given to data structure for model estimation and interpretation of the results, highlighting methodological comparisons and practical differences. Details related to two statistical software are also provided. Furthermore, reliability modeling could be efficiently applied by practitioners and engineers to solve similar technical problems.


  • Open Access

    A Guideline for Implementing a Robust Optimization of a Complex Multi-Stage Manufacturing Process

    by Francesco Bertocci, Andrea Grandoni, Monica Fidanza, Rossella Berni

    In the industrial production scenario, the goal of engineering is focused on the continuous improvement of the process performance by maximizing the effectiveness of the manufacturing and the quality of the products. In order to address these aims, the advanced robust process optimization techniques have been designed, implemented, and applied to the manufacturing process of ultrasound (US) probes for medical imaging. The suggested guideline plays a key role for improving a complex multi-stage manufacturing process; it consists of statistical methods applied for improving the product quality, and for achieving a higher productivity, jointly with engineering techniques oriented to problem solving. Starting from the Six Sigma approach, the high definition of the production process was analyzed through a risk analysis, and thus providing a successful implementation of the PDCA (plan-do-check-act) methodology. Therefore, the multidisciplinary analysis is carried out by applying statistical models and by detecting the latent failures by means of NDT (non-destructive testing), i.e., scanning acoustic microscopy (SAM). The presented approach, driven by the statistical analysis, allows the engineers to distinguish the potential weak points of the complex manufacturing, in order to implement the corrective actions. Furthermore, in this paper we illustrate this approach by considering a pilot study, e.g., a process of US probes for medical imaging, by detailing all the guideline steps.


  • Open Access

    Scanning Acoustic Microscopy (SAM): A Robust Method for Defect Detection during the Manufacturing Process of Ultrasound Probes for Medical Imaging

    by Francesco Bertocci, Andrea Grandoni, Tatjana Djuric-Rissner

    The main aim of this paper is to provide the feasibility of non-destructive testing (NDT) method, such as scanning acoustic microscopy (SAM), for damage detection in ultrasound (US) probes for medical imaging during the manufacturing process. In a highly competitive and demanding electronics and biomedical market, reliable non-destructive methods for quality control and failure analysis of electronic components within multi-layered structures are strongly required. Any robust non-destructive method should be capable of dealing with the complexity of miniaturized assemblies, such as the acoustic stack of ultrasonic transducers. In this work, the application of SAM in an industrial scenario was studied for 24 samples of a phased array probe, in order to investigate potential internal integrity, to detect damages, and to assess the compliance of high-demanding quality requirements. Delamination, non-homogeneous layers with micron-thickness, and entrapped air bubbles (blisters) in the bulk of US probe acoustic stacks were detected and studied. Analysis of 2D images and defects visualization by means of ultrasound-based NDT method were compared with electroacoustic characterization (also following as pulse-echo test) of the US probe through an ad-hoc measurement system. SAM becomes very useful for defect detection in multilayered structures with a thickness of some microns by assuring low time-consuming (a limit for other NDT techniques) and quantitative analyses based on measurements. The study provides a tangible contribution and identifies an advantage for manufacturers of ultrasound probes that are oriented toward continuous improvement devoted to the process capability, product quality, and in-process inspection.


  • Open Access

    Optimization of Gas Sensors Based on Advanced Nanomaterials through Split-Plot Designs and GLMMs

    by Rossella Berni, Francesco Bertocci

    This paper deals with the planning and modeling of a split-plot experiment to improve novel gas sensing materials based on Perovskite, a nano-structured, semi-conductor material that is sensitive to changes in the concentration of hazardous gas in the ambient air. The study addresses both applied and theoretical issues. More precisely, it focuses on (i) the detection of harmful gases, e.g., NO 2 and CO, which have a great impact on industrial applications as well as a significantly harmful impact on human health; (ii) the planning and modeling of a split-plot design for the two target gases by applying a dual-response modeling approach in which two models, e.g., location and dispersion models, are estimated; and (iii) a robust process optimization conducted in the final modeling step for each target gas and for each gas sensing material, conditioned to the minimization of the working temperature. The dual-response modeling allows us to achieve satisfactory estimates for the process variables and, at the same time, good diagnostic valuations. Optimal solutions are obtained for each gas sensing material while also improving the results achieved from previous studies.

  • A clearance measurement system based on on-component multilayer tri-axial capacitive probe

    by Tommaso Addabbo, Francesco Bertocci, Ada Fort, Marco Mugnaini, Enza Panzardi, Valerio Vignoli, Chiara Cinelli


  • Design of Medical Ultrasound Probe: Measurement System for the Characterization of Reverberations

    by Francesco Bertocci, Lorenzo Francalanci, Ramona De Luca, Michele Bassani, Francesca Gelli, Paolo Palchetti

    A measurement system which characterizes the reverberations occurring in medical ultrasound (US) images has been developed. Reverberations cause multiple copies of anatomical structure interfaces, degrading image quality, accuracy and leading to potential misdiagnosis. The system is suitable for characterizing the reverberations during design phase of US probes and for investigating the root causes of these distortions that influence the medical images. The system, based on non-destructive evaluation of the materials, is implemented by means of an ad hoc measurement chain, in which a silicone membrane specially designed to generate reverberations, replicates the same acoustic properties of human skin. The experimental setup based Time of Flight (ToF) of the echoes allows for detecting the reverberation source in the US probe acoustic stack. The proposed quantitative approach can help to predict US probe performances during design stage becoming very useful during imaging validation. Accurate description of the ultrasonic transmit-receive measurement system is discussed by taking into account different material, geometry and thickness of the layers involved in the US probe. Experimental results show that the implemented system can detect the root cause of reverberations by accurately quantifying the layer source providing imaging distortion.


  • A Mathematical Model for Reverberations in Biomedical Ultrasound Transducers: a case study

    by Ramona De Luca, Michele Bassani, Lorenzo Francalanci, Francesco Bertocci, Francesca Gelli, Paolo Palchetti, Davide Coppini, Leonardo Bocchi

    Ultrasonography is a non-invasive diagnostic technique using ultrasound to produce images of internal body structures in real-time. It is a reliable, robust and reproducible diagnostic imaging modality commonly used in the clinical practice. Nevertheless, artifacts may occur in images, leading to misdiagnosis. This paper aims to characterize the reverberation artifacts that cause multiple copies of anatomic structures interfaces, degrading image quality and accuracy. Analyzing and optimizing the mechanical properties of the ultrasound probe play a crucial role in the reduction of reverberations. A simplified two-dimensional mathematical model was developed to simulate the response of an ultrasound transducer to a phantom specially designed to generate reverberations. Such model was verified by experiments using a dedicated water-tank set-up. The investigation of a linear array transducer showed that the model yields good results: R 2 =0.987 for the correlation between the arrival time of the echoes of simulated and experimental signals, and R 2 =0.945 for the correlation of their intensity, at the central frequency. These preliminary results show that the developed model allows for investigating whether changes in the constitutive characteristics of the individual transducer layers reduce reverberations and helps to predict transducer performance in terms of image quality.


  • Open Access

    Optimization of Perovskite Gas Sensor Performance: Characterization, Measurement and Experimental Design

    by Francesco Bertocci, Ada Fort, Valerio Vignoli, Marco Mugnaini, Rossella Berni

    Eight different types of nanostructured perovskites based on YCoO 3 with different chemical compositions are prepared as gas sensor materials, and they are studied with two target gases NO 2 and CO. Moreover, a statistical approach is adopted to optimize their performance. The innovative contribution is carried out through a split-plot design planning and modeling, also involving random effects, for studying Metal Oxide Semiconductors (MOX) sensors in a robust design context. The statistical results prove the validity of the proposed approach; in fact, for each material type, the variation of the electrical resistance achieves a satisfactory optimized value conditional to the working temperature and by controlling for the gas concentration variability. Just to mention some results, the sensing material YCo 0 . 9 Pd 0 . 1 O 3 (Mt1) achieved excellent solutions during the optimization procedure. In particular, Mt1 resulted in being useful and feasible for the detection of both gases, with optimal response equal to +10.23% and working temperature at 312 ∘ C for CO (284 ppm, from design) and response equal to −14.17% at 185 ∘ C for NO 2 (16 ppm, from design). Analogously, for NO 2 (16 ppm, from design), the material type YCo 0 . 9 O 2 . 85 + 1 % Pd (Mt8) allows for optimizing the response value at − 15 . 39 % with a working temperature at 181 . 0 ∘ C, whereas for YCo 0 . 95 Pd 0 . 05 O 3 (Mt3), the best response value is achieved at − 15 . 40 % with the temperature equal to 204 ∘ C.


  • Gas sensing properties of YMnO3 based materials for the detection of NOx and CO

    by Tommaso Addabbo, Francesco Bertocci, Ada Fort, Michele Gregorkiewitz, Marco Mugnaini, Roberto Spinicci, Valerio Vignoli


  • Fast nano-grained Fe3O4 gas sensor for the control of CO and NO2 concentrations

    by Francesco Bertocci, Ada Fort, Marco Mugnaini, Valerio Vignoli, Elisa Bertolucci, Mirko Marracci, Anna Maria Raspolli Galletti, Bernardo Tellini

    In this paper, a resistive-metal-oxide gas sensor based on Fe3O4 nano-powders is presented. The sensor is proposed for the detection of the combustion by-products CO and NO2. The device performance is assessed as a function of both gas concentration and temperature. The developed device has a fast response that makes it suitable for combustion control applications.