Clinical wear of 63 ultrahigh molecular weight polyethylene acetabular components: Effect of starting resin and forming method
Authors:
- 1 January 1999
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 48 (3) , 374-384
Abstract
This study compares the clinical wear rates and implant characteristics of 63 surgically retrieved acetabular components. All components were made by the same manufacturer, implanted by the same surgeon, in articulation against the same type of femoral component, and revised for the same reason; 19 were made from directly compression molded, calcium stearate free, ultrahigh molecular weight polyethylene (UHMWPE) and 44 were made from machined, ram extruded, calcium stearate containing UHMWPE. There were significant differences in wear, duration, and wear rate between the molded (type I) and machined (type II and III) components. Most importantly, the wear rates of type I (molded) components were significantly (p < 0.0001) lower than the wear rates of type II, type III, and type II and III components as a group (all machined). The machined components had wear rates 2.3 times greater than the molded components. The wear rates between the two different groups of machined components (type II and III) were not significantly different. The scanning electron microscope observations did not reveal any major differences in wear mechanisms between the three types of components, although the machined components did show more evidence of brittleness. The molded components were better consolidated (or had less fusion defects) than the ram extruded components. © 1999 John Wiley & Sons, Inc. J Biomed Mater Res (Appl Biomater) 48: 374–384, 1999Keywords
This publication has 7 references indexed in Scilit:
- Total hip arthroplasty with an uncemented femoral component: Excellent results at 10-year follow-upThe Journal of Bone and Joint Surgery. British volume, 1997
- Morphology of rod stock and compression-moulded sheets of ultra-high-molecular-weight polyethylene used in orthopaedic implantsBiomaterials, 1996
- Distinguishing wear and creep in clinically retrieved polyethylene insertsJournal of Biomedical Materials Research, 1995
- Der Einfluß des Verarbeitungsverfahrens auf die Kriech- und Abriebeigenschaften von UHMWPE - The Influence of the Manufacturing Process on Creep and Wear Properties of UHMWPEBiomedical Engineering / Biomedizinische Technik, 1995
- Ultra-high molecular weight polyethylene. The material and its use in total joint implants.Journal of Bone and Joint Surgery, 1994
- Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prosthesesJournal of Biomedical Materials Research, 1983
- Rate of Wear in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975