Eddy current-induced artifacts correction in high gradient strength diffusion MRI with dynamic field monitoring: demonstration in ex vivo human brain imaging
Preprint
Abstract
Purpose To demonstrate the advantages of spatiotemporal magnetic field monitoring to correct eddy current-induced artifacts (ghosting and geometric distortions) in high gradient strength diffusion MRI (dMRI). Methods A dynamic field camera with 16 NMR field probes was used to characterize eddy current fields induced from diffusion gradients for different gradients strengths (up to 300 mT/m), diffusion directions, and shots in a 3D multi-shot EPI sequence on a 3T Connectom scanner. The efficacy of dynamic field monitoring-based image reconstruction was demonstrated on high-resolution whole brain ex vivo dMRI. A 3D multi-shot image reconstruction framework was informed with the actual nonlinear phase evolution measured with the dynamic field camera, thereby accounting for high-order eddy currents fields on top of the image encoding gradients in the image formation model. Results Eddy current fields from diffusion gradients at high gradient strength in a 3T Connectom scanner are highly nonlinear in space and time, inducing high-order spatial phase modulations between odd/even echoes and shots that are not static during the readout. Superior reduction of ghosting and geometric distortion was achieved with dynamic field monitoring compared to ghosting approaches such as navigator- and structured low-rank-based methods or MUSE, followed by image-based distortion correction with eddy. Improved dMRI analysis is demonstrated with diffusion tensor imaging and high-angular resolution diffusion imaging. Conclusion Strong eddy current artifacts characteristic of high gradient strength dMRI can be well corrected with dynamic field monitoring-based image reconstruction, unlike the two-step approach consisting of ghosting correction followed by geometric distortion reduction with eddy.Keywords
This publication has 81 references indexed in Scilit:
- Diffusion imaging of whole, post-mortem human brains on a clinical MRI scannerNeuroImage, 2011
- An ex vivo imaging pipeline for producing high‐quality and high‐resolution diffusion‐weighted imaging datasetsHuman Brain Mapping, 2011
- High resolution diffusion-weighted imaging in fixed human brain using diffusion-weighted steady state free precessionNeuroImage, 2009
- Concomitant field terms for asymmetric gradient coils: Consequences for diffusion, flow, and echo‐planar imagingMagnetic Resonance in Medicine, 2008
- Artifacts and pitfalls in diffusion MRIJournal of Magnetic Resonance Imaging, 2006
- Use of spherical harmonic deconvolution methods to compensate for nonlinear gradient effects on MRI imagesMagnetic Resonance in Medicine, 2004
- Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field gradients in NMR systemsJournal of Magnetic Resonance (1969), 1990
- Optimization of eddy-current compensationJournal of Magnetic Resonance (1969), 1990
- Exact temporal eddy current compensation in magnetic resonance imaging systemsIEEE Transactions on Medical Imaging, 1988
- Least Squares Methods for Ill-Posed Problems with a Prescribed BoundSIAM Journal on Mathematical Analysis, 1970