Effects of calcium dobesilate (CaD) interference on serum creatinine measurements: a national External Quality Assessment (EQA)-based educational survey of drug-laboratory test interactions
- 12 July 2020
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 59 (1) , 139-145
Abstract
Objectives: Drug-laboratory test interactions (DLTIs) are one of the major sources of laboratory errors. Calcium dobesilate (CaD) interference on serum creatinine testing is a widespread problem that has long been ignored in China. A national EQA-based survey was launched to investigate the current status of CaD interference on creatinine routine methods used in China and enhance the education of CaD interference in clinical laboratories. Methods: A descriptive survey was developed to characterize the status quo of Chinese laboratory professionals’ cognition to CaD interference. Four of survey samples which were spiked with/without interference additive were shipped to 175 participant laboratories. The target reference values from a reference measurement procedure were compared against the results from participating laboratories to evaluate the CaD interference on serum creatinine measurements using enzymatic method or Jaffé method. Results: The lack of knowledge of DLTIs and the barriers to collect information from pharmacological and laboratory data systems had become the main problems on implementing DLTIs education in China. A significant negative influence of CaD on enzymatic method was observed regardless of measurement platforms. Jaffé method was generally free from interaction with CaD but showed poor precision and accuracy at low creatinine concentrations. Conclusions: More efforts should be made to enhance the education of DLTIs in clinical laboratories in China.Keywords
Funding Information
- China Capital Health Research and Development of Special Project (2016-2-4017)
This publication has 21 references indexed in Scilit:
- Strong Negative Interference by Calcium Dobesilate in Sarcosine Oxidase Assays for Serum Creatinine Involving the Trinder ReactionMedicine, 2015
- Clinical evaluation of analytical variations in serum creatinine measurements: why laboratories should abandon Jaffe techniquesBMC Nephrology, 2012
- Specificity Characteristics of 7 Commercial Creatinine Measurement Procedures by Enzymatic and Jaffe Method PrinciplesClinical Chemistry, 2012
- A multicentric evaluation of IDMS-traceable creatinine enzymatic assaysClinica Chimica Acta, 2011
- Safety of Calcium Dobesilate in Chronic Venous Disease, Diabetic Retinopathy and HaemorrhoidsDrug Safety, 2004
- A rise in plasma creatinine that is not a sign of renal failure: which drugs can be responsible?Journal of Internal Medicine, 1999
- Multicentre evaluationClinical Chemistry and Laboratory Medicine (CCLM), 1986
- Computerized interpretation of laboratory tests: an overview of systems, basic principles and logic techniquesClinical Biochemistry, 1983
- Use of a clinical laboratory computer to warn of possible drug interference with test resultsComputer Programs in Biomedicine, 1978
- Automated monitoring of drug‐test interactionsClinical Pharmacology & Therapeutics, 1978