A New Electrochemical Sensor Based on Task-Specific Ionic Liquids-Modified Palm Shell Activated Carbon for the Determination of Mercury in Water Samples
Open Access
- 21 July 2014
- Vol. 14 (7) , 13102-13113
Abstract
In this study, a potentiometric sensor composed of palm shell activated carbon modified with trioctylmethylammonium thiosalicylate (TOMATS) was used for the potentiometric determination of mercury ions in water samples. The proposed potentiometric sensor has good operating characteristics towards Hg (II), including a relatively high selectivity; a Nernstian response to Hg (II) ions in a concentration range of 1.0 × 10−9 to 1.0 × 10−2 M, with a detection limit of 1 × 10−10 M and a slope of 44.08 ± 1.0 mV/decade; and a fast response time (~5 s). No significant changes in electrode potential were observed when the pH was varied over the range of 3–9. Additionally, the proposed electrode was characterized by good selectivity towards Hg (II) and no significant interferences from other cationic or anionic species.Keywords
Funding Information
- High Impact Research Grant (UM.C/HIR/MOHE/ENG/43.)
This publication has 42 references indexed in Scilit:
- Determination of Hg in water and wastewater samples by CV-AAS following on-line preconcentration with silver trapJournal of Analytical Chemistry, 2010
- Determination of mercury by electrochemical cold vapor generation atomic fluorescence spectrometry using polyaniline modified graphite electrode as cathodeSpectrochimica Acta Part B: Atomic Spectroscopy, 2010
- Trace mercury determination in drinking and natural water samples by room temperature ionic liquid based-preconcentration and flow injection-cold vapor atomic absorption spectrometryJournal of Hazardous Materials, 2009
- On-line preconcentration and determination of mercury in biological and environmental samples by cold vapor-atomic absorption spectrometryJournal of Hazardous Materials, 2007
- Evaluation of slurry preparation procedures for the determination of mercury by axial view inductively coupled plasma optical emission spectrometry using on-line cold vapor generationSpectrochimica Acta Part B: Atomic Spectroscopy, 2005
- Gas chromatography/plasma spectrometry—an important analytical tool for elemental speciation studiesSpectrochimica Acta Part B: Atomic Spectroscopy, 2004
- Sensitive determination of mercury in tap water by cloud point extraction pre-concentration and flow injection-cold vapor-inductively coupled plasma optical emission spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 2002
- Mechanistic studies on the trapping and desorption of volatile hydrides and mercury for their determination by electrothermal vaporization-inductively-coupled plasma mass spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 2002
- Determination of very low levels of dissolved mercury(II) and methylmercury in river waters by continuous flow with on-line UV decomposition and cold-vapor atomic fluorescence spectrometry after pre-concentration on a silica gel-2-mercaptobenzimidazol sorbentTalanta, 2001
- Simultaneous determination of traces of iron, cobalt, nickel, copper, mercury and lead in water by energy-dispersive x-ray fluorescence spectrometry after preconcentration as their piperazino-1,4-bis(dithiocarbamate) complexesAnalytica Chimica Acta, 1993