Patient-specific uncertainty calibration of deep learning-based autosegmentation networks for adaptive MRI-guided lung radiotherapy
Open Access
- 16 May 2025
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (10) , 105018
Abstract
Objective. Uncertainty assessment of deep learning autosegmentation (DLAS) models can support contour corrections in adaptive radiotherapy (ART), e.g. by utilizing Monte Carlo Dropout (MCD) uncertainty maps. However, poorly calibrated uncertainties at the patient level often render these clinically nonviable. We evaluated population-based and patient-specific DLAS accuracy and uncertainty calibration and propose a patient-specific post-training uncertainty calibration method for DLAS in ART. Approach. The study included 122 lung cancer patients treated with a low-field MR-linac (80/19/23 training/validation/test cases). Ten single-label 3D-U-Net population-based baseline models (BM) were trained with dropout using planning MRIs (pMRIs) and contours for nine organs-at-riks (OARs) and gross tumor volumes (GTVs). Patient-specific models (PS) were created by fine-tuning BMs with each test patient’s pMRI. Model uncertainty was assessed with MCD, averaged into probability maps. Uncertainty calibration was evaluated with reliability diagrams and expected calibration error (ECE). A proposed post-training calibration method rescaled MCD probabilities for fraction images in BM (calBM) and PS (calPS) after fitting reliability diagrams from pMRIs. All models were evaluated on fraction images using Dice similarity coefficient (DSC), 95th percentile Hausdorff distance (HD95) and ECE. Metrics were compared among models for all OARs combined (n = 163), and the GTV (n = 23), using Friedman and posthoc-Nemenyi tests (α = 0.05). Main results. For the OARs, patient-specific fine-tuning significantly (p < 0.001) increased median DSC from 0.78 (BM) to 0.86 (PS) and reduced HD95 from 14 mm (BM) to 6.0 mm (PS). Uncertainty calibration achieved substantial reductions in ECE, from 0.25 (BM) to 0.091 (calBM) and 0.22 (PS) to 0.11 (calPS) (p < 0.001), without significantly affecting DSC or HD95 (p > 0.05). For the GTV, BM performance was poor (DSC = 0.05) but significantly (p < 0.001) improved with PS training (DSC = 0.75) while uncertainty calibration reduced ECE from 0.22 (PS) to 0.15 (calPS) (p = 0.45). Significance. Post-training uncertainty calibration yields geometrically accurate DLAS models with well-calibrated uncertainty estimates, crucial for ART applications.Keywords
Funding Information
- Deutsche Krebshilfe (70115909)
- Munich Medical & Clinician Scientist Program
This publication has 39 references indexed in Scilit:
- Personalized deep learning auto‐segmentation models for adaptive fractionated magnetic resonance‐guided radiation therapy of the abdomenMedical Physics, 2024
- CloverNet – Leveraging Planning Annotations for Enhanced Procedural MR Segmentation: An Application to Adaptive Radiation Therapy2024
- Deep learning for autosegmentation for radiotherapy treatment planning: State-of-the-art and novel perspectivesStrahlentherapie und Onkologie, 2024
- Patient‐specific transfer learning for auto‐segmentation in adaptive 0.35 T MRgRT of prostate cancer: a bi‐centric evaluationMedical Physics, 2022
- Personalized auto‐segmentation for magnetic resonance imaging–guided adaptive radiotherapy of prostate cancerMedical Physics, 2022
- Varian ethos online adaptive radiotherapy for prostate cancer: Early results of contouring accuracy, treatment plan quality, and treatment timeJournal of Applied Clinical Medical Physics, 2021
- Prospects for daily online adaptive radiotherapy via ethos for prostate cancer patients without nodal involvement using unedited CBCT auto‐segmentationJournal of Applied Clinical Medical Physics, 2021
- Assessing Reliability and Challenges of Uncertainty Estimations for Medical Image Segmentation2019
- Left-Ventricle Quantification Using Residual U-Net2019
- U-Net: Convolutional Networks for Biomedical Image Segmentation2015