Eye lens dosimetry: task 2 within the ORAMED project
Open Access
- 20 February 2011
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 144 (1-4) , 473-477
Abstract
The ORAMED (Optimization of RAdiation protection for MEDical staff) project is funded by EU-EURATOM within the 7° Framework Programme. Task 2 of the project is devoted to study the dose to the eye lens. The study was subdivided into various topics, starting from a critical revision of the operational quantity Hp(3), with the corresponding proposal of a cylindrical phantom simulating as best as possible the head in which the eyes are located, the production of a complete set of air kerma to dose equivalent conversion coefficients for photons from 10 keV to 10 MeV, and finally, the optimisation of the design of a personal dosemeter well suited to respond in terms of Hp(3). The paper presents some preliminary results.Keywords
This publication has 5 references indexed in Scilit:
- Risk of Cataract after Exposure to Low Doses of Ionizing Radiation: A 20-Year Prospective Cohort Study among US Radiologic TechnologistsAmerican Journal of Epidemiology, 2008
- Postoperative Cataract Cases among Atomic Bomb Survivors: Radiation Dose Response and ThresholdRadiation Research, 2007
- Dosimetry for a Study of Low-Dose Radiation Cataracts among Chernobyl Clean-up WorkersRadiation Research, 2007
- Cataracts among Chernobyl Clean-up Workers: Implications Regarding Permissible Eye ExposuresRadiation Research, 2007
- Thermoluminescence Efficiency of LiF:Mg,Cu,P (MCP-N) Detectors to Photons, Beta-Electrons, Alpha Particles and Thermal NeutronsRadiation Protection Dosimetry, 1994