Abstract
Diffuse correlation spectroscopy (DCS) is a non-invasive optical method for the measurement of blood flow (BF).1 In DCS, the tissue is illuminated by a long coherence length near-infrared laser, and the speckle pattern formed by moving scatterers, mostly red blood cells, modulates the detected light. The decay of the measured temporal intensity autocorrelation function [g2(τ)] originated by the speckle fluctuations provides an index of blood flow (BFi),2 with units cm2/s. To maximize the contrast of the measured speckle, single-mode fibers are used, greatly limiting potential photon throughput. Current DCS devices employing single-photon avalanche photodiodes (SPAD) detectors and laser sources at 700 to 850 nm typically operate at a source-detector (SD) separation of 25 mm and an acquisition rate of 1 Hz.3 Larger SD separations are desirable for improving brain sensitivity and reducing scalp signal contamination, especially in the adult population. Faster acquisition rates are needed to detect fast BF dynamics and effectively remove the large pulsatile systemic component from the cerebral signals. Unfortunately, the low signal-to-noise ratio (SNR) of current devices limits the acquisition rates and prevents the use of SD separations >2.5  cm.