Experimental helium‐beam radiography with a high‐energy beam: Water‐equivalent thickness calibration and first image‐quality results
Authors:
Open Access
- 20 June 2022
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 49 (8) , 5347-5362
Abstract
Purpose: A clinical implementation of ion‐beam radiography (iRad) is envisaged to provide a method for on‐couch verification of ion‐beam treatment plans. The aim of this work is to introduce and evaluate a method for quantitative water‐equivalent thickness (WET) measurements for a specific helium‐ion imaging system for WETs that are relevant for imaging thicker body parts in the future.Methods: Helium‐beam radiographs (αRads) are measured at the Heidelberg Ion‐beam Therapy Center with an initial beam energy of 239.5 MeV/u. An imaging system based on three pairs of thin silicon pixel detectors is used for ion path reconstruction and measuring the energy deposition (dE) of each particle behind the object to be imaged. The dE behind homogeneous plastic blocks is related to their well‐known WETs between 280.6 and 312.6 mm with a calibration curve that is created by a fit to measured data points. The quality of the quantitative WET measurements is determined by the uncertainty of the measured WET of a single ion (single‐ion WET precision) and the deviation of a measured WET value to the well‐known WET (WET accuracy). Subsequently, the fitted calibration curve is applied to an energy deposition radiograph of a phantom with a complex geometry. The spatial resolution (modulation transfer function at 10 % —MTF10%) and WET accuracy (mean absolute percentage difference—MAPD) of the WET map are determined.Results: In the optimal imaging WET‐range from ∼280 to 300 mm, the fitted calibration curve reached a mean single‐ion WET precision of 1.55 0.00%. Applying the calibration to an ion radiograph (iRad) of a more complex WET distribution, the spatial resolution was determined to be MTF10% = 0.49 0.03 lp/mm and the WET accuracy was assessed as MAPD to 0.21 %.Conclusions: Using a beam energy of 239.5 MeV/u and the proposed calibration procedure, quantitative αRads of WETs between ∼280 and 300 mm can be measured and show high potential for clinical use. The proposed approach with the resulting image qualities encourages further investigation toward the clinical application of helium‐beam radiography.Keywords
This publication has 62 references indexed in Scilit:
- Characterizing the modulation transfer function (MTF) of proton/carbon radiography using Monte Carlo simulationsMedical Physics, 2013
- Comprehensive analysis of proton range uncertainties related to patient stopping-power-ratio estimation using the stoichiometric calibrationPhysics in Medicine & Biology, 2012
- Range uncertainties in proton therapy and the role of Monte Carlo simulationsPhysics in Medicine & Biology, 2012
- Stopping power for particle therapy: The generic library libdEdx and clinically relevant stopping-power ratios for light ionsInternational Journal of Radiation Biology, 2011
- FITPix — fast interface for Timepix pixel detectorsJournal of Instrumentation, 2011
- Proton range verification using a range probe: definition of concept and initial analysisPhysics in Medicine & Biology, 2010
- Deriving the modulation transfer function of CT from extremely noisy edge profilesRadiological Physics and Technology, 2009
- A maximum likelihood proton path formalism for application in proton computed tomographyMedical Physics, 2008
- An analytical approximation of the Bragg curve for therapeutic proton beamsMedical Physics, 1997
- Instrumentation for treatment of cancer using proton and light-ion beamsReview of Scientific Instruments, 1993