Ambient lighting alters pattern electroretinogram P50 peak time and spatial sensitivity
Open Access
- 14 August 2024
- journal article
- research article
- Published by Springer Nature in Documenta Ophthalmologica
- Vol. 149 (2) , 77-86
Abstract
Purpose: Our aim was to explore the effect of ambient lighting on the pattern ERG (PERG). Methods: We compared PERGs recorded in two conditions; room lights on and room lights off. PERGs from 21 adult participants were recorded from each eye to high contrast checks of 50’ side width, reversing 3rps in a large (30°) and then standard (15°) field. This was performed first in lights-ON conditions, then 2 min after the room lights were switched off. A minimum of 2 averages of 300 trials were acquired for each condition. A subset of 10 participants had PERGs recorded to a 50’ check width with a range of stimulus contrasts (96–18%), also to a range of different check widths (100’–12’) at high contrast in both ambient lighting conditions in a 30° field. Results: The lights-ON P50 median peak time (PT) was 3 ms earlier than the lights-OFF P50 from the 30° field (range 0–5 ms) and 15° field (range 0–6 ms). The earlier lights-ON P50 PT was evident at different stimulus contrasts, even after accounting for stimulus contrast reductions associated with stray ambient lighting in lights-ON conditions. Lights-OFF and lights-ON P50 PT were similar to different check widths; the lights-OFF P50 PT to a 50’ check width matched the lights-ON P50 PT to a 25’ check width. Conclusion: PERG P50 PT in lights-ON ambient light conditions can be earlier than in lights-OFF ambient light conditions. The difference in P50 PT with ambient light may reflect alterations in spatial sensitivity associated with retinal adaptation. These results emphasise the clinical importance of consistent ambient lighting for PERG recording and calibration.Keywords
This publication has 36 references indexed in Scilit:
- ISCEV standard for clinical pattern electroretinography (2024 update)Documenta Ophthalmologica, 2024
- ISCEV guidelines for calibration and verification of stimuli and recording instruments (2023 update)Documenta Ophthalmologica, 2023
- ISCEV guide to visual electrodiagnostic proceduresDocumenta Ophthalmologica, 2018
- ISCEV standard for clinical pattern electroretinography (PERG): 2012 updateDocumenta Ophthalmologica, 2012
- Light- and dark-adapted electroretinograms (ERGs) and ocular pigmentation: comparison of brown- and blue-eyed cohortsDocumenta Ophthalmologica, 2010
- The contrast characteristic of the pattern electroretinogram depends on temporal frequencyGraefe's Archive for Clinical and Experimental Ophthalmology, 1999
- Check size tuning of the pattern electroretingoram: a reappraisalDocumenta Ophthalmologica, 1996
- Retinal receptive fields under different adaptation levels studied with pattern-evoked ERGGraefe's Archive for Clinical and Experimental Ophthalmology, 1987
- The spatial organization of retinal receptive fields in light and darkness as revealed by the pattern electroretinogramDocumenta Ophthalmologica, 1986
- A chart demonstrating variations in acuity with retinal positionVision Research, 1974