Comparison of Automated Platelet Counts and Potential Effect on Transfusion Decisions in Cancer Patients
Open Access
- 1 November 2013
- journal article
- research article
- Published by Oxford University Press (OUP) in American Journal of Clinical Pathology
- Vol. 140 (5) , 747-754
Abstract
Objectives: To evaluate the accuracy of platelet counts from various hematology analyzers using a reference immunologic method. Methods: We tested 403 samples with platelet counts less than 50 × 109/L with the Advia (Siemens, Tarrytown, NY), Sysmex (Mundelein, IL), and Abbott (Santa Clara, CA) analyzers. Results: All methods showed a positive bias, especially at less than 20 × 109/L and less than 10 × 109/L. Undertransfusion risk ranged from 9.1% to 43.3 % in the groups below 20 × 109/L and below 10 × 109/L, respectively. For patients with optical counts more than 10 × 109/L and CD61 less than 10 × 109/L, 64.5% were transfused within 24 hours of the reported count, while 35.5% were transfused in more than 24 hours, after a subsequent optical platelet count of 10 × 109/L or less was reported. Conclusions: Although optical and impedance methods were shown to be falsely increased in severely thrombocytopenic samples, further studies are needed to determine if more accurate methods would be clinically useful.Keywords
This publication has 30 references indexed in Scilit:
- Therapeutic platelet transfusion versus routine prophylactic transfusion in patients with haematological malignancies: an open-label, multicentre, randomised studyThe Lancet, 2012
- Dose of Prophylactic Platelet Transfusions and Prevention of HemorrhageThe New England journal of medicine, 2010
- Platelet count evaluation using three automated haematology analysers compared with the immunoplatelet reference method, and estimation of possible inadequate platelet transfusionInternational Journal of Laboratory Hematology, 2009
- Impact of immunological platelet counting (by the platelet/RBC ratio) on haematological practiceCytometry Part B: Clinical Cytometry, 2005
- The ADVIA 2120 Hematology System: Flow Cytometry-Based Analysis of Blood and Body Fluids in the Routine Hematology LaboratoryLaboratory Hematology, 2005
- Performance Evaluation of the ADVIA 2120 Hematology Analyzer: An International Multicenter Clinical TrialLaboratory Hematology, 2005
- An Interlaboratory Study of a Candidate Reference Method for Platelet CountingAmerican Journal of Clinical Pathology, 2001
- Improved Platelet Counting Using Two-Dimensional Laser Light ScatterAmerican Journal of Clinical Pathology, 2000
- A Rapid and Accurate Closed-Tube Immunoassay for Platelets on an Automated Hematology AnalyzerAmerican Journal of Clinical Pathology, 2000
- Safety of stringent prophylactic platelet transfusion policy for patients with acute leukaemiaThe Lancet, 1991