Stopping power and range estimations in proton therapy based on prompt gamma timing: motion models and automated parameter optimization
Authors:
- Julius WernerJWJulius Werner
- Francesco PennazioFPFrancesco Pennazio
- Niklas SchmidNSNiklas Schmid
- Elisa FiorinaEFElisa Fiorina
- Davide BersaniDBDavide Bersani
- Piergiorgio CerelloPCPiergiorgio Cerello
- Jona KasprzakJKJona Kasprzak
- Nicola MoscoNMNicola Mosco
- Sahar RanjbarSRSahar Ranjbar
- Roberto SacchiRoberto Sacchi
- Veronica FerreroVFVeronica Ferrero
- Magdalena RafecasMRMagdalena Rafecas
Open Access
- 15 July 2024
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 69 (14) , 14NT02
Abstract
Objective. Particle therapy treatments are currently limited by uncertainties of the delivered dose. Verification techniques like Prompt-Gamma-Timing-based Stopping Power Estimation (PGT-SPE) may allow for reduction of safety margins in treatment planning. Approach. From Prompt-Gamma-Timing measurements, we reconstruct the spatiotemporal distribution of prompt gamma emissions, which is linked to the average motion of the primary particles. The stopping power is determined by fitting a model of the average particle motion. Here, we compare a previously published implementation of the particle motion model with an alternative formulation and present two formulations to automatically select the hyperparameters of our procedure. The performance was assessed using Monte-Carlo simulations of proton beams (60 MeV–219 MeV) impinging on a homogeneous PMMA phantom. Main results. The range was successfully determined within a standard deviation of 3 mm for proton beam energies from 70 MeV to 219 MeV. Stopping power estimates showed errors below 5% for beam energies above 160 MeV. At lower energies, the estimation performance degraded to unsatisfactory levels due to the short range of the protons. The new motion model improved the estimation performance by up to 5% for beam energies from 100 MeV to 150 MeV with mean errors ranging from 6% to 18%. The automated hyperparameter optimization matched the average error of previously reported manual selections, while significantly reducing the outliers. Significance. The data-driven hyperparameter optimization allowed for a reproducible and fast evaluation of our method. The updated motion model and evaluation at new beam energies bring us closer to applying PGT-SPE in more complex scenarios. Direct comparison of stopping power estimates between treatment planning and measurements during irradiation would offer a more direct verification than other secondary-particle-based techniques.Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (Cluster of Excellence EXC 2167 (Project no. 39088)
- Deutscher Akademischer Austauschdienst (mobility fellowship within the IFI programme)
- Commissione Scientifica Nazionale 5, Instituto Nazionale di Fisica Nucleare (Young Investigators Grant MERLINO no. 23246/21)
This publication has 17 references indexed in Scilit:
- Calibration method and performance of a time‐of‐flight detector to measure absolute beam energy in proton therapyMedical Physics, 2023
- Consensus guide on CT-based prediction of stopping-power ratio using a Hounsfield look-up table for proton therapyRadiotherapy and Oncology, 2023
- The MERLINO project:characterization of LaBr3:Ce detectors for stopping power monitoring in proton therapyJournal of Instrumentation, 2022
- Proton therapy monitoring: spatiotemporal emission reconstruction with prompt gamma timing and implementation with PET detectorsPhysics in Medicine & Biology, 2022
- A time-of-flight-based reconstruction for real-time prompt-gamma imaging in proton therapyPhysics in Medicine & Biology, 2021
- Processing of prompt gamma-ray timing data for proton range measurements at a clinical beam deliveryPhysics in Medicine & Biology, 2019
- Range assessment in particle therapy based on promptγ-ray timing measurementsPhysics in Medicine & Biology, 2014
- The FLUKA Code: Developments and Challenges for High Energy and Medical ApplicationsNuclear Data Sheets, 2014
- Range uncertainties in proton therapy and the role of Monte Carlo simulationsPhysics in Medicine & Biology, 2012
- Properties and Mitigation of Edge Artifacts in PSF-Based PET ReconstructionIEEE Transactions on Nuclear Science, 2011