Validation of a Body-Conducted Sound Sensor for Respiratory Sound Monitoring and a Comparison with Several Sensors
Open Access
Abstract
The ideal respiratory sound sensor exhibits high sensitivity, wide-band frequency characteristics, and excellent anti-noise properties. We investigated the body-conducted sound sensor (BCS) and verified its usefulness in respiratory sound monitoring through comparison with an air-coupled microphone (ACM) and acceleration sensor (B & K: 8001). We conducted four experiments for comparison: (1) estimation by equivalent circuit model of sensors and measurement by a sensitivity evaluation system; (2) measurement of tissue-borne sensitivity-to-air-noise sensitivity ratio (SRTA); (3) respiratory sound measurement through a simulator; and (4) actual respiratory sound measurement using human subjects. For (1), the simulation and measured values of all the sensors showed good agreement; BCS demonstrated sensitivity ~10 dB higher than ACM and higher sensitivity in the high-frequency segments compared with 8001. In (2), BCS showed high SRTA in the 600–1000 and 1200–2000-Hz frequency segments. In (3), BCS detected wheezes in the high-frequency segments of the respiratory sound. Finally, in (4), the sensors showed similar characteristics and features in the high-frequency segments as the simulators, where typical breathing sound detection was possible. BCS displayed a higher sensitivity and anti-noise property in high-frequency segments compared with the other sensors and is a useful respiratory sound sensor.Keywords
This publication has 35 references indexed in Scilit:
- Using Body-Conducted Acoustic Sensors for Human-Robot Communication in Noisy EnvironmentsInternational Journal of Advanced Robotic Systems, 2013
- Technology for Enhancing Chest Auscultation in Clinical SimulationRespiratory Care, 2011
- Silent-speech enhancement using body-conducted vocal-tract resonance signalsSpeech Communication, 2010
- Frequency characteristics of several non-audible murmur (NAM) microphonesAcoustical Science and Technology, 2009
- Comparison of lung sound transducers using a bioacoustic transducer testing systemJournal of Applied Physiology, 2006
- Design, construction, and evaluation of a bioacoustic transducer testing (BATT) system for respiratory soundsIEEE Transactions on Biomedical Engineering, 2006
- A Robust Method for Estimating Respiratory Flow Using Tracheal Sounds EntropyIEEE Transactions on Biomedical Engineering, 2006
- Non-Audible Murmur (NAM) RecognitionIEICE Transactions on Information and Systems, 2006
- Spectral characteristics of chest wall breath sounds in normal subjects.Thorax, 1995
- Abnormal Lung Sounds in Patients With Asthma During Episodes With Normal Lung FunctionChest, 1994