Continuous-Flow Stable Sulfur Isotope Analysis of Organic and Inorganic Compounds by EA-MC-ICPMS
Open Access
- 13 May 2024
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 96 (21) , 8510-8517
Abstract
Elemental analysis (EA) coupled to isotope ratio mass spectrometry (IRMS) is a well-established method to derive stable isotope ratios of sulfur (34S/32S). Conversion of sulfur to SO2 by EA and measurement of SO2 isotopologues by IRMS represents the simplest and most versatile method to accomplish isotope measurement of sulfur even in bulk samples. Yet, interferences by oxygen isotopes in SO2 often impair the precision and trueness of measured results. In the current study, we coupled EA to multicollector inductively coupled plasma mass spectrometry (MC-ICPMS) to establish a method that avoids such interferences due to direct measurement of S+ ions. In addition, measurement of the 33S/32S isotope ratios is possible, thus representing the first bulk method that is suitable to study mass-independent isotope fractionation (MIF). Analytical precision (σ) of available Ag2S and BaSO4 reference materials (RMs) was, on average, 0.2 mUr for δ33S and δ34S, never exceeding 0.3 mUr within this study (1 mUr = 1‰ = 0.001). Measured δ34S values of reference materials agreed within ±0.2 mUr of officially reported values. Measurement of wood samples yielded good precision (0.2 mUr) for sulfur amounts as low as 3.5 μg, but precision deteriorated for samples at lower sulfur contents due to poor peak shape. Finally, we explored cross-calibration of organic liquids separated via gas chromatography (GC) against solid RMs combusted via EA that avoids challenging offline conversion of RMs. Results indicate good precision (≤0.08 mUr) and acceptable trueness (≤0.34 mUr) for determination of δ34S, demonstrating the future potential of such an approach.Keywords
This publication has 38 references indexed in Scilit:
- A comprehensive evaluation of sulfur isotopic analysis (δ34S and δ33S) using multi-collector ICP-MS with characterization of reference materials of geological and biological originAnalytica Chimica Acta, 2022
- Improved analysis of micro‐ and nanomole‐scale sulfur multi‐isotope compositions by gas source isotope ratio mass spectrometryRapid Communications in Mass Spectrometry, 2016
- Revised δ 34 S reference values for IAEA sulfur isotope reference materials S‐2 and S‐3Rapid Communications in Mass Spectrometry, 2009
- Inorganic Mass Spectrometry2007
- Coupled N, C and S stable isotope measurements using a dual‐column gas chromatography systemRapid Communications in Mass Spectrometry, 2007
- Signal improvement in elemental analyzer/continuous flow isotope ratio mass spectrometry for samples with low sulfur contents using a pre‐concentration technique for on‐line concentration adjustmentRapid Communications in Mass Spectrometry, 2006
- Measurement of sulfur isotope composition ( δ 34 S) by multiple‐collector thermal ionization mass spectrometry using a 33 S‐ 36 S double spikeRapid Communications in Mass Spectrometry, 2005
- Oxygen isotope corrections for online δ 34 S analysisRapid Communications in Mass Spectrometry, 2002
- On the determination of chlorine and sulfur in wood and woodbased panelsEuropean Journal of Wood and Wood Products, 1984
- A 33S Anomaly in the allende meteoriteGeochimica et Cosmochimica Acta, 1977