Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion
Authors:
- Michael W. KissickMKMichael W. Kissick
- Sarah A. BoswellSBSarah A. Boswell
- Robert JerajRJRobert Jeraj
- T. Rockwell MackieTMT. Rockwell Mackie
- 23 June 2005
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 32 (7Part1) , 2346-2350
Abstract
The interplay between a constant scan speed and intrafraction oscillatory motion produces interesting fluence intensity modulations along the axis of motion that are sensitive to the motion function, as originally shown in a classic paper by Yu et al. [Phys. Med. Biol. 43, 91-104 (1998)]. The fluence intensity profiles are explored in this note for an intuitive understanding, then compared with Yu et al., and finally further explored for the effects of low scan speed and random components of both intrafraction and interfraction motion. At slow scan speeds typical of helical tomotherapy, these fluence intensity modulations are only a few percent. With the addition of only a small amount of cycle-to-cycle randomness in frequency and amplitude, the fluence intensity profiles change dramatically. It is further shown that after a typical 30-fraction treatment, the sensitivities displayed in the single fraction fluence intensity profiles greatly diminish.Keywords
Funding Information
- National Institutes of Health (T32 CA09206, PO1 CA88960)
This publication has 8 references indexed in Scilit:
- Influence of intra-fractional breathing movement in step-and-shoot IMRTPhysics in Medicine & Biology, 2004
- Effects of motion on the total dose distributionSeminars in Radiation Oncology, 2004
- The effects of intra‐fraction organ motion on the delivery of intensity‐modulated field with a multileaf collimatorMedical Physics, 2003
- An experimental investigation on intra-fractional organ motion effects in lung IMRT treatmentsPhysics in Medicine & Biology, 2003
- A method for the reconstruction of four‐dimensional synchronized CT scans acquired during free breathingMedical Physics, 2003
- Motion adaptive x-ray therapy: a feasibility studyPhysics in Medicine & Biology, 2000
- The effects of intra-fraction organ motion on the delivery of dynamic intensity modulationPhysics in Medicine & Biology, 1998
- Tomotherapy: A new concept for the delivery of dynamic conformal radiotherapyMedical Physics, 1993