Virtual Monochromatic Images from Dual-Energy Multidetector CT: Variance in CT Numbers from the Same Lesion between Single-Source Projection-based and Dual-Source Image-based Implementations

Authors:

Abstract
The results of our phantom investigation suggest that single-source projection-based and dual-source image-based dual energy hardware implementations produce significant differences in virtual monochromatic CT numbers from the same iodine-containing lesion. PurposeTo determine the variance in virtual monochromatic computed tomography (CT) numbers from the same lesion, comparing the two clinically available dual-energy multidetector CT hardware implementations (single-source projection-based and dual-source image-based), in a phantom-based simulated abdominal environment.Materials and MethodsThis phantom-based study was exempt from institutional review board oversight. Polyethylene terephthalate spheres (15 and 18 mm) with two iodine-to-saline dilutions (0.8 and 1.2 mg of iodine per millilliter) were serially suspended in a cylindrical polypropylene bottle filled with diluted iodinated contrast material. The bottle was placed into a 36-cm-wide torso-shaped water phantom simulating the abdomen of a medium-sized patient. Dual-energy (80/140 kVp) and single-energy (100 and 120 kVp) scans were obtained with single-source and dual-source multidetector CT implementations. Virtual monochromatic images were reconstructed at energy levels of 40–140 keV (in 10-keV incr...

This publication has 14 references indexed in Scilit: