Virtual Monochromatic Images from Dual-Energy Multidetector CT: Variance in CT Numbers from the Same Lesion between Single-Source Projection-based and Dual-Source Image-based Implementations
Authors:
- 1 April 2016
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 279 (1) , 269-277
Abstract
The results of our phantom investigation suggest that single-source projection-based and dual-source image-based dual energy hardware implementations produce significant differences in virtual monochromatic CT numbers from the same iodine-containing lesion. PurposeTo determine the variance in virtual monochromatic computed tomography (CT) numbers from the same lesion, comparing the two clinically available dual-energy multidetector CT hardware implementations (single-source projection-based and dual-source image-based), in a phantom-based simulated abdominal environment.Materials and MethodsThis phantom-based study was exempt from institutional review board oversight. Polyethylene terephthalate spheres (15 and 18 mm) with two iodine-to-saline dilutions (0.8 and 1.2 mg of iodine per millilliter) were serially suspended in a cylindrical polypropylene bottle filled with diluted iodinated contrast material. The bottle was placed into a 36-cm-wide torso-shaped water phantom simulating the abdomen of a medium-sized patient. Dual-energy (80/140 kVp) and single-energy (100 and 120 kVp) scans were obtained with single-source and dual-source multidetector CT implementations. Virtual monochromatic images were reconstructed at energy levels of 40–140 keV (in 10-keV incr...Keywords
This publication has 14 references indexed in Scilit:
- Maximizing Iodine Contrast-to-Noise Ratios in Abdominal CT Imaging through Use of Energy Domain Noise Reduction and Virtual Monoenergetic Dual-Energy CTRadiology, 2015
- Dual-Energy MDCT in Hypervascular Liver Tumors: Effect of Body Size on Selection of the Optimal Monochromatic Energy LevelAmerican Journal of Roentgenology, 2014
- Increasing the Precision of CT Measurements with Dual-Energy ScanningRadiology, 2014
- Dual-Energy Liver CT: Effect of Monochromatic Imaging on Lesion Detection, Conspicuity, and Contrast-to-Noise Ratio of Hypervascular Lesions on Late Arterial PhaseAmerican Journal of Roentgenology, 2014
- Impact of Dual-Energy Multi–Detector Row CT with Virtual Monochromatic Imaging on Renal Cyst Pseudoenhancement: In Vitro and in Vivo StudyRadiology, 2014
- Dual-Energy Multi–Detector Row CT with Virtual Monochromatic Imaging for Improving Patient-to-Patient Uniformity of Aortic Enhancement during CT Angiography: An in Vitro and in Vivo StudyRadiology, 2014
- Usefulness of the Virtual Monochromatic Image in Dual-Energy Spectral CT for Decreasing Renal Cyst Pseudoenhancement: A Phantom StudyAmerican Journal of Roentgenology, 2012
- Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image qualityMedical Physics, 2011
- Virtual Monochromatic Spectral Imaging with Fast Kilovoltage Switching: Improved Image Quality as Compared with That Obtained with Conventional 120-kVp CTRadiology, 2011
- Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kVp switching dual energy CT scannerMedical Physics, 2011