Temperature Correction to Enhance Blood Glucose Monitoring Accuracy Using Electrical Impedance Spectroscopy
Open Access
Abstract
Electrical methods are among the primarily studied non-invasive glucose measurement techniques; however, various factors affect the accuracy of the sensors used. Of these, the temperature is a critical factor; hence, the effects of temperature on the electrical properties of blood components are investigated in this study. Furthermore, the changes in the electrical properties of blood according to the glucose level are corrected by considering the effects of temperature on the electrical properties. An impedance sensor is developed and used to measure whole blood impedance in 10 healthy participants at various temperatures and glucose levels. Subsequently, the conductivities of the plasma and cytoplasm were extracted. Changes in the electrical properties of the blood components are then analyzed using linear regression and repeated measures ANOVA. The electrical conductivities of plasma and cytoplasm increased with increasing temperatures (plasma: 0.0397 (slope), 0.7814 (R2), cytoplasm: 0.014 (slope), 0.694 (R2)). At three values of increasing glucose levels (85.4, 158.1, and 271.8 mg/dL), the electrical conductivities of the plasma and cytoplasm decreased. These tendencies are more significant upon temperature corrections (p-values; plasma: 0.001, 0.001, cytoplasm: 0.003, 0.002). The relationships between temperature and electrical conductivity changes can thus be used for temperature corrections in blood glucose measurement.Keywords
Funding Information
- Samsung Advanced Institute of Technology (GI06070)
This publication has 33 references indexed in Scilit:
- Review of Non-Invasive Glucose Sensing Techniques: Optical, Electrical and Breath AcetoneSensors, 2020
- Benefits and limitations of continuous glucose monitoring in type 1 diabetesExpert Review of Endocrinology & Metabolism, 2020
- Diabetes Fact Sheets in Korea, 2018: An Appraisal of Current StatusDiabetes & Metabolism Journal, 2019
- Temperature-Corrected Fluidic Glucose Sensor Based on Microwave ResonatorSensors, 2018
- Non-invasive glucose monitoring technology in diabetes management: A reviewAnalytica Chimica Acta, 2012
- Blood Glucose Measurement Principles of Non-invasive Blood Glucose Meter: Focused on the Detection Methods of Blood GlucoseJournal of Biomedical Engineering Research, 2012
- Characteristics of a multisensor system for non invasive glucose monitoring with external validation and prospective evaluationBiosensors and Bioelectronics, 2011
- Cutaneous Blood Perfusion as a Perturbing Factor for Noninvasive Glucose MonitoringDiabetes Technology & Therapeutics, 2010
- Continuous Glucose MonitoringDiabetes Care, 2005
- Impact of Posture and Fixation Technique on Impedance Spectroscopy Used for Continuous and Noninvasive Glucose MonitoringDiabetes Technology & Therapeutics, 2004