Sampling size in the verification of manufactured-supplied air kerma strengths
Authors:
- Luis Isaac RamosLRLuis Isaac Ramos
- Rafael Martinez MongeRMRafael Martinez Monge
- 18 October 2005
- journal article
- Published by Wiley in Medical Physics
- Vol. 32 (11) , 3375-3378
Abstract
Quality control mandate that the air kerma strengths (S(K)) of permanent seeds be verified, this is usually done by statistics inferred from 10% of the seeds. The goal of this paper is to proposed a new sampling method in which the number of seeds to be measured will be set beforehand according to an a priori statistical level of uncertainty. The results are based on the assumption that the S(K) has a normal distribution. To demonstrate this, the S(K) of each of the seeds measured was corrected to ensure that the average S(K) of its sample remained the same. In this process 2030 results were collected and analyzed using a normal plot. In our opinion, the number of seeds sampled should be determined beforehand according to an a priori level of statistical uncertainty.Keywords
This publication has 2 references indexed in Scilit: