A five-channel, diamond photoconducting x-ray detector array for z-pinch experiments
Author:
- 1 October 1992
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 63 (10) , 5056-5058
Abstract
We have built a five‐channel, x‐ray detector array based on diamond photoconducting detectors (PCDs). The diamond elements have dimensions of 3 mm × 1 mm × 1 mm (or 0.5 mm). We use diamond PCDs for their stability, flat spectral response, and low leakage currents. The good time response of diamond PCDs is due to the 100‐ps electron/hole recombination time. Filters were designed to give information in the 1–10‐keV spectral region. Calibration of the diamond PCDs showed sensitivities between 4 and 7 × 10−4 A/W for a bias of 100 V. We shall present data from z‐pinch experiments on Saturn.Keywords
This publication has 6 references indexed in Scilit:
- Soft x-ray detection with diamond photoconductive detectorsReview of Scientific Instruments, 1990
- Photoconducting x-ray detectors for Z-pinch experimentsReview of Scientific Instruments, 1988
- Flat response detectors for the vacuum ultraviolet and soft x-ray region: InP:Fe photoconductorsJournal of Applied Physics, 1986
- A technique for i n s i t u calibration of an x-ray streak camera in the nanosecond regime using a high density Z-pinchJournal of Applied Physics, 1985
- High-speed ultraviolet and x-ray-sensitive InP photoconductive detectorsApplied Physics Letters, 1982
- Pinch of a Gas JetPhysical Review Letters, 1978