Direct measurement of absorbed dose to water in HDR I192r brachytherapy: Water calorimetry, ionization chamber, Gafchromic film, and TG-43
Authors:
- Iwan KawrakowIKIwan Kawrakow
- Jan SeuntjensJSJan Seuntjens
- 31 March 2010
- journal article
- Published by Wiley in Medical Physics
- Vol. 37 (4) , 1924-1932
Abstract
Gafchromic film and ionometric calibration procedures for HDR 192Ir brachytherapy sources in terms of dose rate to water are presented and the experimental results are compared to the TG-43 protocol as well as with the absolute dose measurement results from a water calorimetry-based primary standard. EBT-1 Gafchromic films, an A1SL Exradin miniature Shonka thimble type chamber, and an SI HDR 1000 Plus well-type chamber (Standard Imaging, Inc., Middleton, WI) with an ADCL traceable Sk calibration coefficient (following the AAPM TG-43 protocol) were used. The Farmer chamber and Gafchromic film measurements were performed directly in water. All results were compared to direct and absolute absorbed dose to water measurements from a 4 degrees C stagnant water calorimeter. Based on water calorimetry, the authors measured the dose rate to water to be 361 +/- 7 microGy/(h U) at a 55 mm source-to-detector separation. The dose rate normalized to air-kerma strength for all the techniques agree with the water calorimetry results to within 0.83%. The overall 1-sigma uncertainty on water calorimetry, ionization chamber, Gafchromic film, and TG-43 dose rate measurement amounts to 1.90%, 1.44%, 1.78%, and 2.50%, respectively. This work allows us to build a more realistic uncertainty estimate for absorbed dose to water determination using the TG-43 protocol. Furthermore, it provides the framework necessary for a shift from indirect HDR 192Ir brachytherapy dosimetry to a more accurate, direct, and absolute measurement of absorbed dose to water.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (RGPIN 298181)
- Canadian Institutes of Health Research
This publication has 40 references indexed in Scilit:
- On the accuracy of techniques for obtaining the calibration coefficient NK of Ir192 HDR brachytherapy sourcesMedical Physics, 2006
- Determination of the reference air kerma rate for brachytherapy sources and the related uncertaintyMedical Physics, 2004
- Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculationsMedical Physics, 2004
- Monte Carlo‐aided dosimetry of a new high dose‐rate brachytherapy sourceMedical Physics, 1998
- Monte Carlo and TLD dosimetry of an high dose‐rate brachytherapy sourceMedical Physics, 1998
- TLD, diode and Monte Carlo dosimetry of an192Ir source for high dose-rate brachytherapyPhysics in Medicine & Biology, 1995
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Absorbed dose distribution measurements in brachytherapy using ferrous sulphate gel and magnetic resonance imagingBritish Journal of Radiology, 1994
- Imaging of HDR brachytherapy dose distributions using NMR Fricke-gelatin dosimetryMagnetic Resonance Imaging, 1994
- Calibration of 192Ir high‐dose‐rate afterloading systemsMedical Physics, 1991