Assessment of volume and leak measurements during CPAP using a neonatal lung model
Authors:
- H S FischerHFH S Fischer
- C C RoehrCRC C Roehr
- H ProquittéHPH Proquitté
- R R WauerRWR R Wauer
- G SchmalischGSG Schmalisch
- 1 January 2008
- journal article
- research article
- Published by IOP Publishing in Physiological Measurement
- Vol. 29 (1) , 95-107
Abstract
Although several commercial devices are available which allow tidal volume and air leak monitoring during continuous positive airway pressure (CPAP) in neonates, little is known about their measurement accuracy and about the influence of air leaks on volume measurement. The aim of this in vitro study was the validation of volume and leak measurement under CPAP using a commercial ventilatory device, taking into consideration the clinical conditions in neonatology. The measurement accuracy of the Leoni ventilator (Heinen & Löwenstein, Germany) was investigated both in a leak-free system and with leaks simulated using calibration syringes (2-10 ml, 20-100 ml) and a mechanical lung model. Open tubes of variable lengths were connected for leak simulation. Leak flow was measured with the flow-through technique. In a leak-free system the mean relative volume error +/-SD was 3.5 +/- 2.6% (2-10 ml) and 5.9 +/- 0.7% (20-60 ml), respectively. The influence of CPAP level, driving flow, respiratory rate and humidification of the breathing gas on the volume error was negligible. However, an increasing F(i)O(2) caused the measured tidal volume to increase by up to 25% (F(i)O(2) = 1.0). The relative error +/- SD of the leak measurements was -0.2 +/- 11.9%. For leaks > 19%, measured tidal volume was underestimated by more than 10%. In conclusion, the present in vitro study showed that the Leoni allowed accurate volume monitoring under CPAP conditions similar to neonates. Air leaks of up to 90% of patient flow were reliably detected. For an F(i)O(2) > 0.4 and for leaks > 19%, a numerical correction of the displayed volume should be performed.Keywords
This publication has 41 references indexed in Scilit:
- Mouth closing device (chinstrap) reduces mouth leak during nasal CPAPSleep Medicine, 2004
- Changes in lung volume and work of breathing: A comparison of two variable‐flow nasal continuous positive airway pressure devices in very low birth weight infantsPediatric Pulmonology, 2003
- Are tidal volume measurements in neonatal pressure‐controlled ventilation accurate?Pediatric Pulmonology, 2002
- Erfassung des Maskenlecks unter CPAP bei Patienten mit obstruktivem Schlafapnoesyndrom1Pneumologie, 2000
- Randomised controlled trial of respiratory system compliance measurements in mechanically ventilated neonatesArchives of Disease in Childhood - Fetal and Neonatal Edition, 1998
- Effect of endotracheal tube leakage on functional residual capacity determination by nitrogen washout method in a small‐sized lung modelPediatric Pulmonology, 1994
- Accurate, noninvasive quantitation of expiratory gas leak from uncuffed infant endotracheal tubesPediatric Pulmonology, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Continuous positive airway pressure selectively reduces obstructive apnea in preterm infantsThe Journal of Pediatrics, 1985
- Effects of Continuous Distending Pressure on Lung Volumes and Lung Mechanics in the Immediate Neonatal PeriodNeonatology, 1977