Temperature Increase in Human Cadaver Retina During Direct Illumination by Femtosecond Laser Pulses
Authors:
- Eric MikulaEMEric Mikula
- Ronald M. KurtzRKRonald M. Kurtz
- Tibor JuhaszTJTibor Juhasz
- 1 April 2010
- journal article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 26 (4) , 272-277
Abstract
PURPOSE: Femtosecond lasers have been approved by the US Food and Drug Administration for ophthalmic surgery, including use in creating corneal flaps in LASIK surgery. During normal operation, approximately 50% to 60% of laser energy may pass beyond the cornea, with potential effects on the retina. As a model for retinal laser exposure during femtosecond corneal surgery, we measured the temperature rise in human cadaver retinas during direct illumination by the laser.METHODS: The temperature increase induced by a 150-kHz iFS Advanced Femtosecond Laser (Abbott Medical Optics) in human cadaver retinas was measured in situ using an infrared thermal imaging camera. To model the geometry of the eye during the surgery, an approximate 11×11-mm excised section of human cadaver retina was placed 17 mm behind the focus of the laser beam. The temperature field was observed in 10 cadaver retina samples at energy levels ranging from 0.4 to 1.6 µJ (corresponding approximately to surgical energies of 0.8 to 3.2 µJ per pulse).RESULTS: Maximal temperature increases up to 1.15°C (corresponding to 3.2 µJ and 52-second illumination) were observed in the cadaver retina sections with little variation in temperature profiles between specimens for the same laser energy illumination.CONCLUSIONS: The commercial iFS Advanced Femtosecond Laser operating with pulse energies at approximately the lower limit of the range evaluated in this study would be expected to result in a 0.2°C temperature increase and do not therefore present a safety hazard to the retina. [J Refract Surg. 2010;26:272–277.]Keywords
This publication has 28 references indexed in Scilit:
- Characterization of submicrojoule femtosecond laser corneal tissue dissectionJournal of Cataract & Refractive Surgery, 2008
- The use of the femtosecond laser in the customization of corneal flaps in laser in situ keratomileusisCurrent Opinion in Ophthalmology, 2007
- Comparison of the corneal response to laser in situ keratomileusis with flap creation using the FS15 and FS30 femtosecond lasersJournal of Cataract & Refractive Surgery, 2007
- Understanding heat transfer mechanisms during the cooling phase of blow molding using infrared thermographyNDT & E International, 2005
- Mini-invasive corneal surgery and imaging with femtosecond lasersOptics Express, 2004
- Monte Carlo method in the mathematical simulation of the laser-induced transscleral hyperthermia of intraocular neoplasmsJournal of Optical Technology, 2000
- Features of the absorption of optical radiation by ocular tissues and by liquids replacing the vitreous humorJournal of Optical Technology, 1999
- Corneal refractive surgery with femtosecond lasersIEEE Journal of Selected Topics in Quantum Electronics, 1999
- Non-thermal ablation of neural tissue with femtosecond laser pulsesApplied Physics B, 1998
- Retinal-Temperature Increases Produced by Intense Light Sources*Journal of the Optical Society of America, 1970