On the robustness of multilateration of ionoacoustic signals for localization of the Bragg peak at pre-clinical proton beam energies in water
Open Access
- 8 May 2023
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 68 (10) , 105010
Abstract
Objectives. The energy deposited in a medium by a pulsed proton beam results in the emission of thermoacoustic waves, also called ionoacoustics (IA). The proton beam stopping position (Bragg peak) can be retrieved from a time-of-flight analysis (ToF) of IA signals acquired at different sensor locations (multilateration). This work aimed to assess the robustness of multilateration methods in proton beams at pre-clinical energies for the development of a small animal irradiator. Approach. The accuracy of multilateration performed using different algorithms; namely, time of arrival and time difference of arrival, was investigated in-silico for ideal point sources in the presence of realistic uncertainties on the ToF estimation and ionoacoustic signals generated by a 20 MeV pulsed proton beam stopped in a homogeneous water phantom. The localisation accuracy was further investigated experimentally based on two different measurements with pulsed monoenergetic proton beams at energies of 20 and 22 MeV. Main results. It was found that the localisation accuracy mainly depends on the position of the acoustic detectors relative to the proton beam due to spatial variation of the error on the ToF estimation. By optimally positioning the sensors to reduce the ToF error, the Bragg peak could be located in-silico with an accuracy better than 90 μm (2% error). Localisation errors going up to 1 mm were observed experimentally due to inaccurate knowledge of the sensor positions and noisy ionoacoustic signals. Significance. This study gives a first overview of the implementation of different multilateration methods for ionoacoustics-based Bragg peak localisation in two- and three-dimensions at pre-clinical energies. Different sources of uncertainty were investigated, and their impact on the localisation accuracy was quantified in-silico and experimentally.Keywords
Funding Information
- European Research Council (SIRMIO project, Grant agreement No. 725539)
This publication has 40 references indexed in Scilit:
- Robust optimization in intensity-modulated proton therapy to account for anatomy changes in lung cancer patientsRadiotherapy and Oncology, 2015
- Ionoacoustic characterization of the proton Bragg peak with submillimeter accuracyMedical Physics, 2015
- Proton beam characterization by proton-induced acoustic emission: simulation studiesPhysics in Medicine & Biology, 2014
- The FLUKA Code: Developments and Challenges for High Energy and Medical ApplicationsNuclear Data Sheets, 2014
- Physical and biological factors determining the effective proton rangeMedical Physics, 2013
- In vivoproton range verification: a reviewPhysics in Medicine & Biology, 2013
- Range uncertainties in proton therapy and the role of Monte Carlo simulationsPhysics in Medicine & Biology, 2012
- k-Wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fieldsJournal of Biomedical Optics, 2010
- Empirical relationships between acoustic parameters in human soft tissuesAcoustics Research Letters Online, 2000
- The generalized correlation method for estimation of time delayIEEE Transactions on Acoustics, Speech, and Signal Processing, 1976