Spectral calibration of photon‐counting detectors at high photon flux
Authors:
Open Access
- 31 August 2022
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 49 (10) , 6368-6383
Abstract
Background Calibration of photon-counting detectors (PCDs) is necessary for quantitatively accurate spectral computed tomography (CT), but the calibration process can be complicated by nonlinear flux-dependent physical factors such as pulse pile-up. Purpose This work develops a method for spectral sensitivity calibration of a PCD-based spectral CT system that incorporates nonlinear flux dependence and can thus be employed at high photon flux. Methods A calibration model for the spectral response and polynomial flux dependence is proposed, which incorporates prior x-ray source spectrum and PCD models and that has a small set of parameters for adjusting to the spectral CT system of interest. The model parameters are determined by fitting transmission data from a known object of known composition: a step-wedge phantom composed of different thicknesses of aluminum, a bone equivalent, and polymethyl methacrylate (PMMA), a soft-tissue equivalent. This fitting employs Tikhonov regularization, and the regularization strength and the polynomial order for the intensity modeling are determined by bias and variance analysis. The spectral calibration and nonlinear intensity correction is validated on transmission measurements through a third material, Teflon, at different x-ray photon flux levels. Results The nonlinear intensity dependence is determined to be accurately accounted for with a third-order polynomial. The calibrated spectral CT model accurately predicts Teflon transmission to within 1% for flux levels up to 50% of the detector maximum. Conclusions The proposed PCD calibration method enables accurate physical modeling necessary for quantitative imaging in spectral CT. Furthermore, the model applies to high flux settings so that acquisition times will not be limited by restricting the spectral CT system to low flux levels.Keywords
Funding Information
- National Institutes of Health (R01‐EB023968, R01‐EB026282, R21‐CA263660)
This publication has 20 references indexed in Scilit:
- A neural network-based method for spectral distortion correction in photon counting x-ray CTPhysics in Medicine & Biology, 2016
- X-ray spectrum estimation from transmission measurements by an exponential of a polynomial model2016
- A cascaded model of spectral distortions due to spectral response effects and pulse pileup effects in a photon‐counting x‐ray detector for CTMedical Physics, 2014
- Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image qualityMedical Physics, 2011
- Material separation in x‐ray CT with energy resolved photon‐counting detectorsMedical Physics, 2011
- CT scanner x‐ray spectrum estimation from transmission measurementsMedical Physics, 2011
- An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x‐ray detectorsMedical Physics, 2010
- Quantitative imaging of element composition and mass fraction using dual‐energy CT: Three‐material decompositionMedical Physics, 2009
- Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomographyPhysics in Medicine & Biology, 2008
- A robust method of x-ray source spectrum estimation from transmission measurements: Demonstrated on computer simulated, scatter-free transmission dataJournal of Applied Physics, 2005