Comparison of dosimetric impact of intra‐fractional setup discrepancy between multiple‐ and single‐isocenter approaches in linac‐based stereotactic radiotherapy of multiple brain metastases
Open Access
- 20 December 2021
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 23 (2) , e13484
Abstract
Introduction: Treatment of multiple brain metastases by linac‐based stereotactic radiotherapy (SRT) can employ either a multiple‐isocenter (MI) or single‐isocenter (SI) approach. The purposes of this study were to evaluate the dosimetric results of MI and SI approaches and compare the impacts of intra‐fractional setup discrepancies on the robustness of respective approaches using isocenter shifts, whether the same magnitude of translational and rotational effects could lead to a significant difference between the two approaches.Methods: Twenty‐two patients with multiple brain metastases treated by linac‐based SRT were recruited. Treatment plans were computed with both the MI and SI approaches. For the MI approach, the isocenter was located at the geometric center of each planning target volumes (PTVs), whereas the isocenter of the SI approach was located midway between the PTV centroids. To simulate the intra‐fractional errors, isocenter displacements including translational and rotational shifts were hypothetically applied. Apart from the dosimetric outcomes of the two approaches, the impact of the isocenter shifts on PTVs and organs at risk (OARs) were recorded in terms of the differences (δ) in dose parameters relative to the reference plan and was then compared between the MI and SI approaches.Results: Both MI and SI plans met the plan acceptance criteria. The mean Paddick conformity index (Paddick CI) and Dmax of most OARs between MI and SI plans did not show a significant difference, except that higher doses to the left optic nerve and optic chiasm were found in SI plans (p = 0.03). After the application of the isocenter shifts, δCI increased with an increase in the magnitude of the isocenter shift. When comparing between MI and SI plans, the δCIs were similar (p > 0.05) for all extents of translational shifts, but δCIs were significantly higher in SI plans after application of all rotations particularly ±1.5° and ±2.0° shifts. Despite the result that the majority of δDMax of OARs were higher in the SI plans, only the differences in the left optic nerve and chiasm showed generally consistent significance after both translational ≥±1 mm and rotational shifts of ≥±1.Conclusion: Both MI and SI approaches could produce clinically acceptable plans. However, isocenter shifts brought dosimetric impacts to both MI and SI approaches and the effects increased with the increase of the shift magnitude. Although similar impacts were shown in plans of both approaches after translational isocenter shift, SI plans were relatively more vulnerable than MI plans to rotational shifts.Keywords
This publication has 44 references indexed in Scilit:
- Hypofractionated radiosurgery for intact or resected brain metastases: defining the optimal dose and fractionationRadiation Oncology, 2013
- Treatment of Five or More Brain Metastases With Stereotactic RadiosurgeryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Dosimetric consequences of translational and rotational errors in frame-less image-guided radiosurgeryRadiation Oncology, 2012
- Frameless linac-based stereotactic radiosurgery (SRS) for brain metastases: analysis of patient repositioning using a mask fixation system and clinical outcomesRadiation Oncology, 2011
- Adjuvant Whole-Brain Radiotherapy Versus Observation After Radiosurgery or Surgical Resection of One to Three Cerebral Metastases: Results of the EORTC 22952-26001 StudyJournal of Clinical Oncology, 2011
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical resultsRadiation Oncology, 2010
- Gamma Knife Radiosurgery for Ten or More Brain MetastasesJournal of Korean Neurosurgical Society, 2008
- Errors and margins in radiotherapySeminars in Radiation Oncology, 2004
- Risk analysis of linear accelerator radiosurgeryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1996