Technical Note: Fabricating Cerrobend grids with 3D printing for spatially modulated radiation therapy: A feasibility study
- 8 October 2015
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 42 (11) , 6269-6273
Abstract
Purpose: Grid therapy has promising applications in the radiation treatment of large tumors. However, research and applications of grid therapy are limited by the accessibility of the specialized blocks that produce the grid of pencil‐like radiation beams. In this study, a Cerrobend grid block was fabricated using the 3D printing technique. Methods: A grid block mold was designed with flared tubes which follow the divergence of the beam. The mold was 3D printed using a resin with the working temperature below 230 °C. The melted Cerrobend liquid at 120 °C was cast into the resin mold to yield a block with a thickness of 7.4 cm. At the isocenter plane, the grid had a hexagonal pattern, with each pencil beam diameter of 1.4 cm; the distance between the beam centers was 2.1 cm. Results: The dosimetric properties of the grid block were studied using small field dosimeters: a pinpoint ionization chamber and a stereotactic diode. For a 6 MV photon beam, its valley‐to‐peak ratio was 20% at dmax and 30% at 10 cm depth; the output factor was 84.9% at dmax and 65.1% at 10 cm depth. Conclusions: This study demonstrates that it is feasible to implement 3D printing technique in applying grid therapy in clinic.Keywords
This publication has 25 references indexed in Scilit:
- High dose bystander effects in spatially fractionated radiation therapyCancer Letters, 2015
- Measurement of neutron dose equivalent outside and inside of the treatment vault of GRID therapyMedical Physics, 2013
- Feasibility of a simple method of hybrid collimation for megavoltage grid therapyMedical Physics, 2013
- Application of Spatially Fractionated Radiation (GRID) to Helical Tomotherapy using a Novel TOMOGRID TemplateTechnology in Cancer Research & Treatment, 2013
- High-Dose Spatially Fractionated GRID Radiation Therapy (SFGRT): A Comparison of Treatment Outcomes With Cerrobend vs. MLC SFGRTInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Ionization chamber gradient effects in nonstandard beam configurationsMedical Physics, 2009
- Radiation-induced bystander signalling in cancer therapyNature Reviews Cancer, 2009
- Dosimetric validation of the MCNPX Monte Carlo simulation for radiobiologic studies of megavoltage grid radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Elevated sphingomyelinase activity and ceramide concentration in serum of patients undergoing high dose spatially fractionated radiation treatment: Implications for endothelial apoptosisCancer Biology & Therapy, 2005
- Therapeutic advantage of grid irradiation for large single fractionsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004