Improving the Glucose Meter Error Grid With the Taguchi Loss Function
Open Access
- 29 December 2015
- journal article
- review article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 10 (4) , 967-970
Abstract
Glucose meters often have similar performance when compared by error grid analysis. This is one reason that other statistics such as mean absolute relative deviation (MARD) are used to further differentiate performance. The problem with MARD is that too much information is lost. But additional information is available within the A zone of an error grid by using the Taguchi loss function. Applying the Taguchi loss function gives each glucose meter difference from reference a value ranging from 0 (no error) to 1 (error reaches the A zone limit). Values are averaged over all data which provides an indication of risk of an incorrect medical decision. This allows one to differentiate glucose meter performance for the common case where meters have a high percentage of values in the A zone and no values beyond the B zone. Examples are provided using simulated data.Keywords
This publication has 6 references indexed in Scilit:
- The Surveillance Error GridJournal of Diabetes Science and Technology, 2014
- Technical Aspects of the Parkes Error GridJournal of Diabetes Science and Technology, 2013
- Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator TrialJournal of Diabetes Science and Technology, 2013
- Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation is Part of the AssessmentJournal of Diabetes Science and Technology, 2013
- Wrong Thinking about Glucose StandardsClinical Chemistry, 2010
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987