We report on the measurements of time resolution for double-sided 3D pixel sensors with a single cell of 50 m 50 m and thickness of 285 m, fabricated at IMB-CNM and irradiated with reactor neutrons from 8 1MeV /cm to 1.0 1MeV /cm. The time resolution measurements were conducted using a radioactive source at a temperature of −20 and 20 °C in a bias voltage range of 50–250 V. The reference time was provided by a low gain avalanche detector produced by Hamamatsu. The results are compared to measurements conducted prior to irradiation where a temporal resolution of about 50 ps was measured. These are the first ever timing measurements on an irradiated 3D sensor and which serve as a basis for understanding their performance and to explore the possibility of performing 4D tracking in high radiation environments, such as the innermost tracking layers of future high energy physics experiments.
Keywords
This publication has 7 references indexed in Scilit:
- Properties of HPK UFSD after neutron irradiation up to 6e15 n/cmNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2019
- Timing performance of small cell 3D silicon detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2019
- FCC-hh: The Hadron ColliderThe European Physical Journal Special Topics, 2019
- Radiation hardness of small-pitch 3D pixel sensors up to a fluence of 3×1016 neq/cm2Journal of Instrumentation, 2018
- Beam test results of a 16 ps timing system based on ultra-fast silicon detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2017
- 3D double sided detector fabrication at IMB-CNMNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- 3D — A proposed new architecture for solid-state radiation detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997