Limited (6-h) ambulatory blood pressure monitoring is a valid replacement for the office blood pressure by trained nurse clinician in the diagnosis of hypertension
Authors:
- John W. GravesJGJohn W. Graves
- Carol A. NashCNCarol A. Nash
- Diane E. GrillDGDiane E. Grill
- Kent R. BaileyKBKent R. Bailey
- Sheldon G. ShepsSSSheldon G. Sheps
- 1 August 2005
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 10 (4) , 169-174
Abstract
To assess the ability of limited ambulatory blood pressure monitoring as a valid replacement for office blood pressure measurement done to American Heart Association criteria in diagnosing hypertension. In all, 105 adults, who had been referred for limited ambulatory blood pressure monitoring, participated in the study. Limited ambulatory blood pressure monitoring consisted of 6 h of blood pressure measurement while ambulatory at the Mayo Clinic, using a SpaceLabs 90207 (SpaceLabs Medical, Issaquah, Washington, USA) collecting six readings per hour for the period of observation. The study participants gave consent for three additional consecutive office blood pressure measurements, using a validated aneroid device, done to American Heart Association criteria, by a single hypertension nurse specialist. Mean systolic blood pressure by limited ambulatory blood pressure monitoring was 137.9±14.2 mmHg and for the nurse, 137.9±20.1 mmHg. Mean diastolic blood pressure by limited ambulatory blood pressure monitoring was 81.5±9.7 mmHg and for the nurse, 74.3±11.9 mmHg. The intermethod difference for systolic blood pressure was 0.03±12.5 mmHg and diastolic blood pressure, –7.2±8.0 mmHg. Using <140/90 as criteria factor, limited ambulatory blood pressure monitoring and the trained nurse agreed 77% of the time on whether the patient was hypertensive. This agreement increased to 81% if the participant's referral blood pressure was ≥140/90. Limited ambulatory blood pressure monitoring is an excellent replacement for office blood pressure, done to American Heart Association criteria, in diagnosing hypertension. This avoids issues of variability introduced by the observers, such as digit preference and bias, and increases reproducibility of blood pressure measurements. The appropriate normal value for limited ambulatory blood pressure monitoring is <140/90 mmHg compared with <135/85 mmHg used in 24-h ambulatory blood pressure monitoring.Keywords
This publication has 13 references indexed in Scilit:
- Clinical decision-making in hypertension using an automated (BpTRU™) measurement deviceJournal of Human Hypertension, 2003
- Use of an automated blood pressurerecording device, the BpTRU, to reduce the“white coat effect” in routine practiceAmerican Journal of Hypertension, 2003
- European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurementJournal Of Hypertension, 2003
- Blood pressure measurement - an observational study of 21 public health nursesJournal of Clinical Nursing, 2001
- Are Aneroid Sphygmomanometers Accurate in Hospital and Clinic Settings?Archives of internal medicine (1960), 2001
- Is nurse-measured blood pressure a valid substitute for ambulatory blood pressure monitoring?Blood Pressure Monitoring, 2000
- Human blood pressure determination by sphygmomanometry.Circulation, 1993
- Terminal digit preference, random error, and bias in routine clinical measurement of blood pressureJournal of Clinical Epidemiology, 1993
- Sources of error in recording the blood pressure of patients with hypertension in general practice.BMJ, 1984
- Auscultatory Measurement of Blood Pressure Performed by the Doctor on DutyActa Medica Scandinavica, 1975