A Parallel-Cone Collimator for High-Energy SPECT
Open Access
- 5 February 2015
- journal article
- Published by Society of Nuclear Medicine in Journal of Nuclear Medicine
- Vol. 56 (3) , 476-482
Abstract
In single photon emission computed tomography (SPECT) using high-energy photon emitting isotopes, such as I-131, parallel hole collimators with thick septa are required to limit septal penetration at the cost of sensitivity and resolution. This study investigates a parallel hole collimator with cone shaped holes, which is designed to limit collimator penetration while preserving resolution and sensitivity. The objective is to demonstrate that a single slice prototype of the Parallel Cone (PC) collimator is capable of improving the image quality of high-energy SPECT. Methods: The image quality of the PC collimator was quantitatively compared to that of clinically used Low-Energy High-Resolution (LEHR; for Tc-99m) and High-Energy General-Purpose (HEGP; for I-131 and F-18) parallel hole collimators. First, Monte Carlo simulations of single and double point sources were performed to assess sensitivity and resolution by comparing point spread functions (PSFs). Second, a prototype PC collimator was used in an experimental phantom study to assess and compare contrast recovery coefficients and image noise. Results: Monte Carlo simulations showed reduced broadening of the PSF due to collimator penetration for the PC collimator as compared with the HEGP collimator (e.g. 0.9 cm vs. 1.4 cm FWHM for I-131). Simulated double point sources placed 2 cm apart were separately detectable for the PC collimator, whereas this was not the case for I-131 and F-18 at distances from the collimator face ≥10 cm for the HEGP collimator. The sensitivity, measured over the simulated profiles as the total amount of counts per decay, was found to be higher for the LEHR and HEGP collimators than for the PC collimator (e.g. 3.1x10-5 vs. 2.9x10-5 counts per decay for I-131). However, at equal noise level, phantom measurements showed that contrast recovery coefficients were similar for the PC and LEHR collimators for Tc-99m, but that the PC collimator significantly improved the contrast recovery coefficients as compared with the HEGP collimator for I-131 and F-18. Conclusion: High-energy SPECT imaging with a single slice prototype of the proposed PC collimator has shown the potential for significantly improved image quality in comparison with standard parallel hole collimators.Keywords
This publication has 18 references indexed in Scilit:
- Quantitative Monte Carlo–Based 90Y SPECT ReconstructionJournal of Nuclear Medicine, 2013
- MIRD Pamphlet No. 23: Quantitative SPECT for Patient-Specific 3-Dimensional Dosimetry in Internal Radionuclide TherapyJournal of Nuclear Medicine, 2012
- Radioiodine Scintigraphy with SPECT/CT: An Important Diagnostic Tool for Thyroid Cancer Staging and Risk StratificationJournal of Nuclear Medicine, 2012
- Quantitative Evaluation of Scintillation Camera Imaging Characteristics of Isotopes Used in Liver RadioembolizationPLOS ONE, 2011
- Yttrium-90-labeled microsphere tracking during liver selective internal radiotherapy by bremsstrahlung pinhole SPECT: feasibility study and evaluation in an abdominal phantomEJNMMI Research, 2011
- Unmatched projector/backprojector pair for demultiplexing in multipinhole emission computed tomographyOptical Engineering, 2010
- Small animal imaging with multi-pinhole SPECTMethods, 2009
- Yttrium-90 microsphere radioembolization for the treatment of liver malignancies: a structured meta-analysisEuropean Radiology, 2008
- Nuclear Data Sheets for A = 131Nuclear Data Sheets, 2006
- Performance characteristics of a 511-keV collimator for imaging positron emitters with a standard gamma-cameraEuropean Journal of Nuclear Medicine and Molecular Imaging, 1992