MR-based Thermometry of Laser Induced Thermotherapy: Temperature Accuracy and Temporal Resolution In vitro at 0.2 and 1.5 T Magnetic Field Strengths
- 7 March 2012
- journal article
- basic science
- Published by Wiley in Lasers in Surgery and Medicine
- Vol. 44 (3) , 257-265
Abstract
Purpose To evaluate MR‐thermometry using fast MR sequences for laser induced interstitial thermotherapy (LITT) at 0.2 and 1.5 T systems. Methods & Materials In‐vitro experiments were performed using Agarose gel mixture and lobes of porcine liver. MR‐thermometry was performed by means of longitudinal relaxation time (T1) and proton resonance frequency shift (PRF) methods under acquisition of amplitude and phase shift images. Four different sequences were used for T1 thermometry: A gradient‐echo (GRE), a True Fast Imaging with Steady Precession (TRUFI), a Saturation Recovery Turbo‐FLASH (SRTF), and an Inversion Recovery Turbo‐FLASH (IRTF) sequence (FLASH—Fast Low Angle Shot). PRF was measured with four sequences: Two fast‐spoiled GRE sequences (one as WIP sequence), a Turbo‐FLASH (TFL) sequence (WIP sequence), and a multiecho‐TrueFISP sequence. Temperature was controlled and verified using a fiber‐optic Luxtron device. The temperature was correlated with the MR measurement. Results All sequences showed a good linear correlation R2 = 0.97–0.99 between the measured temperature and the MR‐thermometry measurements. The only exception was the TRUFI sequence in the Agarose phantom that showed a non‐linear calibration curve R2 = 0.39–0.67. At 1.5 T, the Agarose experiments revealed similar temperature accuracies of 4–6°C for all sequences excluding TRUFI. During experiments with the liver, the PRF sequences showed better performance than the T1, with accuracies of 5–12°C, contrary to the T1 sequences at 14–18°C. The accuracy of the Siemens PRF‐FLASH sequence was 5.1°C. At 0.2 T, the Agarose experiments provided the highest accuracy of 3.3°C for PRF measurement. At the liver experiments the T1 sequences SRTF and FLASH revealed the best accuracies at 6.4 and 7.0°C. Conclusion The accuracy and speed of MR temperature measurements are sufficient for controlling the temperature‐based tumor destruction. For 0.2 T systems SRTF and FLASH sequences are recommended. For 1.5 T systems SRTF and FLASH are the most accurate. Lasers Surg. Med. 44:257–265, 2012.Keywords
This publication has 24 references indexed in Scilit:
- Multipathway sequences for MR thermometryMagnetic Resonance in Medicine, 2011
- Absolute MR thermometry using time‐domain analysis of multi‐gradient‐echo magnitude imagesMagnetic Resonance in Medicine, 2010
- Malignant Liver Tumors Treated with MR Imaging–guided Laser-induced Thermotherapy: Experience with Complications in 899 Patients (2,520 lesions)Radiology, 2002
- MR temperature measurement in liver tissue at 0.23 T with a steady-state free precession sequenceMagnetic Resonance in Medicine, 2002
- Quantitative MR temperature monitoring of high-intensity focused ultrasound therapyMagnetic Resonance Imaging, 1999
- Comparison of four magnetic resonance methods for mapping small temperature changesPhysics in Medicine & Biology, 1999
- Monitoring and visualization techniques for MR‐guided laser ablations in an open MR systemJournal of Magnetic Resonance Imaging, 1998
- Temperature dependence of canine brain tissue diffusion coefficient measuredin vivowith magnetic resonance echo-planar imagingInternational Journal of Hyperthermia, 1995
- Verification of a hyperthermia model method using MR thermometryInternational Journal of Hyperthermia, 1995
- Temperature distribution measurements in two‐dimensional NMR imagingMedical Physics, 1983