Design Optimization of an Axial Blood Pump With Computational Fluid Dynamics
Authors:
- 1 March 2008
- journal article
- biomedical engineering
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 54 (2) , 150-155
Abstract
A fully implantable, axial flow blood pump has been developed in our hospital. Both in vitro and in vivo tests showed that the hemolysis and thrombus characteristics of the pump were in an acceptable but not in an ideal range. Computational fluid dynamics (CFD) and in vitro test results showed that the pump worked at off-design point with a low hydraulic efficiency; CFD analysis also showed regions of reverse flow in the diffuser, which not only decreases the pump's hydrodynamic efficiency, but also increases its overall potential for blood trauma and thrombosis. To make a blood pump atraumatic and nonthrombogenic, several methods were taken to reach a final model of the optimized blood pump using CFD, which decreased the rotational speed from 9,000 to 8,000 rpm, and the design flow rate from 11 to 6 L/min. More significantly, the flow separation and recirculation in the diffuser region were eliminated, which mitigated the traumatic and thrombus effect on blood. The acceptable results of the numerical simulations encourage additional in vitro and in vivo studies.Keywords
This publication has 18 references indexed in Scilit:
- Early In Vivo Experience With the Pediatric Jarvik 2000 HeartASAIO Journal, 2007
- Mechanical Circulatory Support as a Bridge to Heart Transplantation: What Remains? Long-term Emotional Sequelae in Patients and SpousesThe Journal of Heart and Lung Transplantation, 2007
- Design Optimization of Blood Shearing Instrument by Computational Fluid DynamicsArtificial Organs, 2005
- Computational Fluid Dynamics Analysis of an Intra‐Cardiac Axial Flow PumpArtificial Organs, 2003
- Computational Fluid Dynamics and Experimental Validation of a Microaxial Blood PumpASAIO Journal, 2001
- Assessment of Hemolysis Related Quantities in a Microaxial Blood Pump by Computational Fluid DynamicsArtificial Organs, 2001
- Development and clinical application of the MicroMed DeBakey VADCurrent Opinion in Cardiology, 2000
- Development of an Implantable Ventricular Assist SystemThe Annals of Thoracic Surgery, 1997
- Model for a General Mechanical Blood Damage PredictionArtificial Organs, 1995
- Hemolysis During Membrane Plasma Separation With Pulsed Flow Filtration EnhancementJournal of Biomechanical Engineering, 1994