In situ trapping of antimony hydride on iridium-coated tungsten coil and interference studies
Authors:
- 4 April 2008
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 23 (7) , 976-980
Abstract
A continuous flow hydride generation manifold coupled to a tungsten coil atomizer for in situ trapping of antimony hydride was described. Iridium coated tungsten coil atomizer can be used for both on-line atomization and trapping while bare tungsten coil can be used only for on-line atomization mode. The interference effects of several hydride forming elements such as As(III), Se(IV), Sn(II) and Te(IV) were compared with a flame heated quartz tube atomizer and tungsten coil atomizer; on-line atomization and in situ trapping modes. The analytical figures of merit for antimony hydride using 60 s trapping period and iridium coated tungsten coil were found as; linear range 0.25–5 ng ml−1, characteristic concentration 0.029 ng ml−1, the limit of detection (3s) 0.028 ng ml−1 and the precision in terms of % RSD was between 1–3. The iridium coating was stable for at least 200 firings without sensitivity loss. The accuracy of the method was tested with certified reference materials Fortified Water TMDA 61 (NWRI), Waste Water EU-L-1 (SCP) and Tea sample GBW 07605. The results were found in good agreement with the certified values at 95% confidence level.Keywords
This publication has 22 references indexed in Scilit:
- Determination of bismuth in metallurgical materials using a quartz tube atomizer with tungsten coil and flow injection-hydride-generation atomic absorption spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 2002
- A quartz tube atomizer with tungsten coil: a new system for vapor atomization in atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 2002
- Tungsten devices in analytical atomic spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 2002
- Investigation of in situ trapping of selenium and arsenic hydrides within a tungsten tube atomiserJournal of Analytical Atomic Spectrometry, 2001
- Tungsten Coil Devices in Atomic Spectrometry: Absorption, Fluorescence, and EmissionAnalytical Sciences, 2001
- Hyphenated vapour generation atomic absorption spectrometric techniquesJournal of Analytical Atomic Spectrometry, 1999
- Atomic spectrometric detection of hydride forming elements following in situ trapping within a graphite furnaceSpectrochimica Acta Part B: Atomic Spectroscopy, 1996
- Horizontal tungsten coil atomizer and integrated absorbance readout for atomic absorption spectrometryAnalytical Chemistry, 1976
- Ultramicro atomic-absorption spectroscopy with a tungsten-filament atom-reservoirTalanta, 1973
- Commercial tungsten filament atomizer for analytical atomic spectrometryAnalytical Chemistry, 1972