This paper describes a microsystem for automated blood sampling from laboratory mice used in pharmacokinetic studies. Intended to be mounted as a “backpack” on a mouse, it uses a microneedle, reservoir, and an actuator to instantaneously prick the animal for a time-point sample, eliminating the need for a tethered catheter with large dead volume. The blood is collected by capillary effect through a 31–33 gauge microneedle (250–210 μm OD) into a ≈1 μL micromachined steel reservoir. The voice coil actuator provides a peak force of ≈300 mN, which amply exceeds the measured piercing force of mouse skin (i.e., 60–85 mN for a 31-gauge needle with 12° bevel). The sampling system was tested in vitro using a mock vessel with adjustable pressure; the reservoir was filled in In vivo tests on live, anesthetized mice resulted in successful collection of blood into the reservoir. The possible integration of the device in microanalytical systems and the device scalability for multisampling are discussed.