Dose and Image Quality Evaluation of a Dedicated Cone-Beam CT System for High-Contrast Neurologic Applications

Authors:

Abstract
OBJECTIVE. The purpose of our study was to evaluate the dose and image quality performance of a dedicated cone-beam CT (CBCT) scanner in comparison with an MDCT scanner.MATERIALS AND METHODS. The conventional dose metric, CT dose index (CTDI), is no longer applicable to CBCT scanners. We propose to use two dose metrics, the volume average dose and the mid plane average dose, to quantify the dose performance in a circular cone-beam scan. Under the condition of equal mid plane average dose, we evaluated the image quality of a CBCT scanner and an MDCT scanner, including high-contrast spatial resolution, low-contrast spatial resolution, noise level, CT number uniformity, and CT number accuracy.RESULTS. For the sinus scanning protocol, the CBCT system had comparable high-contrast resolution and inferior low-contrast resolution to those obtained with the MDCT scanner when the doses were matched (mid plane average dose 9.2 mGy). The CT number uniformity and accuracy were worse on the CBCT scanner. The image arti...

This publication has 25 references indexed in Scilit: