Evaluation of the spatial resolution characteristics of a cone‐beam breast CT scanner
Authors:
- 22 December 2006
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (1) , 275-281
Abstract
The purpose of this study was to examine the spatial resolution of a prototype pendant‐geometry cone‐beam breast computed tomography (CT) system. Modulation transfer functions (MTFs) of the reconstructed image in the coronal ( and ) plane were computed as a function of the cone angle, the radial distance from the axis of rotation, the size of the reconstruction matrix, the back‐projection filter used, and the number of projections acquired for the reconstruction. The results show that the cone angle and size of the reconstruction matrix have minimal impact on the MTF, while the MTF degraded radially from the axis of rotation (from 0.76 at from axis of rotation down to 0.37 at from axis of rotation at ). The Ramp reconstruction filter increases the MTF near the axis of rotation relative to the Shepp‐Logan filter, while an increase in the number of projections from 500 to 1000 increased the MTF near the periphery of the reconstructed image. The MTF in the direction (anterior‐posterior direction) was also evaluated. The ‐direction MTF values tend to be higher when compared to the coronal MTF (0.85 at ), and tend to be very constant throughout the coronal plane direction. The results suggest that an increase in the MTF for the prototype breast CT system is possible by optimizing various scanning and reconstruction parameters.Keywords
This publication has 42 references indexed in Scilit:
- An improved method for flat‐field correction of flat panel x‐ray detectorMedical Physics, 2006
- The x-ray time of flight method for investigation of ghosting in amorphous selenium-based flat panel medical x-ray imagersMedical Physics, 2005
- Flat panel CT detectors for sub-second volumetric scanning2005
- Mercuric iodide medical imagers for low-exposure radiography and fluoroscopy2004
- Multiple-gain-ranging readout method to extend the dynamic range of amorphous silicon flat-panel imagers2004
- Medical imaging with mercuric iodide direct digital radiography flat-panel x-ray detectors2003
- Theoretical framework for a dynamic cone-beam reconstruction algorithm based on a dynamic particle modelPhysics in Medicine & Biology, 2002
- An efficient Fourier method for 3-D radon inversion in exact cone-beam CT reconstructionIEEE Transactions on Medical Imaging, 1998
- Half‐scan cone‐beam X‐ray microtomography formulaScanning, 1994
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984