Abstract
In this paper, a method is proposed to measure the human respiration rate using a mattress sensor. An unsupervised clustering method, combined with a cluster selection algorithm, was used to measure the respiration rate. The volume change of the upper body during the breathing process was indicated by the change of pressure on the mattress sensor and the respiration waveform was estimated by measuring the pressure change. The measured pressure changes were separated into respiratory-related clusters and noise using the clustering method. Finally, a cluster selection algorithm was used to find the combination of clusters that best represented the respiration. The proposed method was compared with other methods. The results showed that the proposed method had a mean Pearson correlation coefficient of the proposed method was 0.8746, which was the highest among all the methods. The difference between these methods was statistically significant (p<0.001). In addition, a regression analysis was performed to measure the accuracy of the respiration rate. The results showed that the proposed method calculated the respiratory rate most accurately.
Funding Information
  • Industrial Technology Innovation Program (Development of Emotional Cognitive and Sympathetic AI Service Technology for Remote
  • Ministry of Trade, Industry and Energy (20012603)

This publication has 23 references indexed in Scilit: