Absorbed-dose-to-water measurement using alanine in ultra-high-pulse-dose-rate electron beams
Open Access
- 14 October 2022
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 67 (20) , 205011
Abstract
Objective . The aim of the presented study is to evaluate the dose response of the PTB’s secondary standard system, which is based on alanine and electron spin resonance (ESR) spectroscopy measurement, in ultra-high-pulse-dose-rate (UHPDR) electron beams. Approach . The alanine dosimeter system was evaluated in the PTB’s UHPDR electron beams (20 MeV) in a range of 0.15–6.2 Gy per pulse. The relationship between the obtained absorbed dose to water per pulse and the in-beamline charge measurement of the electron pulses acquired using an integrating current transformer (ICT) was evaluated. Monte Carlo simulations were used to determine the beam quality conversion and correction factors required to perform alanine dosimetry. Main results . The beam quality conversion factor from the reference quality 60 Co to 20 MeV obtained by Monte Carlo simulation, 1.010(1), was found to be within the standard uncertainty of the consensus value, 1.014(5). The dose-to-water relative standard uncertainty was determined to be 0.68% in PTB’s UHPDR electron beams. Significance . In this investigation, the dose-response of the PTB’s alanine dosimeter system was evaluated in a range of dose per pulse between 0.15 Gy and 6.2 Gy and no evidence of dose-response dependency of the PTB’s secondary standard system based on alanine was observed. The alanine/ESR system was shown to be a precise dosimetry system for evaluating absorbed dose to water in UHPDR electron beams.Keywords
Funding Information
- EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme (18HLT04)
This publication has 32 references indexed in Scilit:
- High dose‐per‐pulse electron beam dosimetry: Usability and dose‐rate independence of EBT3 Gafchromic filmsMedical Physics, 2017
- Experimental Platform for Ultra-high Dose Rate FLASH Irradiation of Small Animals Using a Clinical Linear AcceleratorInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- Evaluation of alanine as a reference dosimeter for therapy level dose comparisons in megavoltage electron beamsMetrologia, 2015
- Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in miceScience Translational Medicine, 2014
- Difference in the relative response of the alanine dosimeter to megavoltage x-ray and electron beamsPhysics in Medicine & Biology, 2013
- Relative response of alanine dosemeters for high-energy electrons determined using a Fricke primary standardPhysics in Medicine & Biology, 2012
- Uncertainties in alanine/ESR dosimetry at the Physikalisch-Technische BundesanstaltPhysics in Medicine & Biology, 2006
- The PTB water calorimeter for the absolute determination of absorbed dose to water in60Co radiationMetrologia, 2006
- Monte Carlo simulation of a typical 60Co therapy sourceMedical Physics, 1999
- Response of alanine dosimeters at very high dose rateApplied Radiation and Isotopes, 1997