Effects of improper source coupling in frequency-domain near-infrared spectroscopy
Authors:
- T E J NoponenTNT E J Noponen
- K KotilahtiKKK Kotilahti
- I NissiläINI Nissilä
- T KajavaTKT Kajava
- P T MeriläinenPMP T Meriläinen
- 30 April 2010
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (10) , 2941-2960
Abstract
Currently, there is no widely used method to assess the reliability of contact between optodes and tissue in near-infrared spectroscopy (NIRS). In this study we observe a high linear dependence (R(2) approximately 0.99) of the logarithmic modulation amplitude (ln(I(AC))), average intensity (ln(I(DC))) and phase (phi) on the source-detector distance (SDD) ranging from approximately 20 to 50 mm on human forehead measurements. The regression of phi is clearly reduced in measurements where light leakage occurs, mainly due to insufficient contact between the source optode and tissue. Utilizing this observation, a novel criterion to detect light leakage is developed. The criterion is applied to study the reliability of hemodynamic responses measured on the human forehead when breathing carbon dioxide-enriched air and during hyperventilation. The contrast of the signals is significantly lower in measurements which were adversely affected by light leakage. Furthermore, such unreliable signals at SDDs >or= 50 mm correlate significantly (for [HbO2] p < 0.01 and for [HbR] p < 0.001) better with the signals measured at SDDs < 20 mm. Using this method, poor contact between the source optode and tissue can be detected and corrected before the actual measurement, which enables us to avoid the acquisition of low contrast cortical signals.Keywords
This publication has 47 references indexed in Scilit:
- Near infrared technology in neuroscience: past, present and futureNeurological Research, 2009
- Noninvasive determination of the optical properties of adult brain: near-infrared spectroscopy approachJournal of Biomedical Optics, 2004
- Can the cerebral metabolic rate of oxygen be estimated with near-infrared spectroscopy?Physics in Medicine & Biology, 2003
- Simultaneous imaging and optode calibration with diffuse optical tomographyOptics Express, 2001
- Phase measurement of light absorption and scatter in human tissueReview of Scientific Instruments, 1998
- Quantification in tissue near–infrared spectroscopyPhilosophical Transactions of the Royal Society B: Biological Sciences, 1997
- Regional cerebral oxygenation during cardiopulmonary bypassPerfusion, 1995
- The theoretical basis for the determination of optical pathlengths in tissue: temporal and frequency analysisPhysics in Medicine & Biology, 1992
- Estimation of optical pathlength through tissue from direct time of flight measurementPhysics in Medicine & Biology, 1988
- Comparison of time-resolved and -unresolved measurements of deoxyhemoglobin in brain.Proceedings of the National Academy of Sciences, 1988