Transmit gain calibration for nonproton MR using the Bloch–Siegert shift

Authors:

Abstract
Transmit gain (B) calibration is necessary for the adjustment of radiofrequency (RF) power levels to the desired flip angles. In proton MRI, this is generally an automated process before the actual scan without any user interaction. For other nuclei, it is usually time consuming and difficult, especially in the case of hyperpolarised MR. In the current work, transmit gain calibration was implemented on the basis of the Bloch–Siegert phase shift. From the same data, the centre frequency, line broadening and SNR could also be determined. The T1 and B0 insensitivity, and the wide range of B over which this technique is effective, make it well suited for nonproton applications. Examples are shown for hyperpolarised 13C and 3He applications. Copyright © 2011 John Wiley & Sons, Ltd.

This publication has 18 references indexed in Scilit: