Automated Exchange Transfusion and Exchange Rate
Authors:
- 1 October 1989
- journal article
- Published by Wiley in Pediatrics International
- Vol. 31 (5) , 572-577
Abstract
An automated blood exchange transfusion (BET) with a two-site technique has been devised by Goldmann et al and by us, using an infusion pump. With this method, we successfully performed exchange transfusions 189 times in the past four years on 110 infants with birth weights ranging from 530 g to 4,000 g. The exchange rate by the automated method was compared with the rate by Diamond's method. Serum bilirubin (SB) levels before and after BET and the maximal SB rebound within 24 hours after BET were: 21.6 +/- 2.4, 11.5 +/- 2.2, and 15.0 +/- 1.5 mg/dl in the automated method, and 22.0 +/- 2.9, 11.2 +/- 2.5, and 17.7 +/- 3.2 mg/dl in Diamond's method, respectively. The result showed that the maximal rebound of the SB level within 24 hours after BET was significantly lower in the automated method than in Diamond's method (p less than 0.01), though SB levels before and after BET were not significantly different between the two methods. The exchange rate was also measured by means of staining the fetal red cells (F cells) both in the automated method and in Diamond's method, and comparing them. The exchange rate of F cells in Diamond's method went down along the theoretical exchange curve proposed by Diamond, while the rate in the automated method was significantly better than in Diamond's, especially in the early stage of BET (p less than 0.01). We believe that the use of this automated method may give better results than Diamond's method in the rate of exchange, because this method is performed with a two-site technique using a peripheral artery and vein.Keywords
This publication has 7 references indexed in Scilit:
- Automated BET, Blood Pressure, and IVHPediatrics International, 1988
- Automated apparatus for exchange transfusionThe Journal of Pediatrics, 1984
- Automated method for exchange transfusion: A new modificationThe Journal of Pediatrics, 1983
- Exchange transfusion in ill newborn infants using peripheral arteries and veinsThe Journal of Pediatrics, 1979
- Percutaneous catheterisation of the radial artery in newborn babies using transillumination.Archives of Disease in Childhood, 1978
- Automated method for exchange transfusion.Archives of Disease in Childhood, 1972
- EXCHANGE-TRANSFUSION APPARATUSThe Lancet, 1966